Braun HNS 11 user manual Operating the nerve stimulator, Switching on / switching off

Page 7

Operating the nerve stimulator

7

 

 

Switching on / switching off

Turn the control knob clockwise out of the O position to switch on the Stimuplex® HNS 11. A brief click signals that this has taken place. At this instant, the device automatically starts a self-test. This includes switching on all segments of the LCD display as a visual check. When the self-test is completed, the nerve stimulator immediately switches to its normal operating mode. The LEDs on the button panel indicate the operating conditions selected (see below). To conserve power, these extinguish after 8 seconds. They light up again immediately when a button is pressed, but this does not change the operating mode. The operating mode can only be changed when the displays are lit.

To switch off the nerve stimulator, turn the control knob anticlockwise to position O.

Selecting the pulse frequency

The selected setting is displayed and stored. You can change the setting by pressing the "Hz" button when the LED display is lit. The Stimuplex® HNS 11 nerve stimulator offers the facility of selecting a frequency of 1 Hz or 2 Hz (one or two pulses per second).

Selecting the pulse width

The pulse width can be set by pressing the ”ms” button. It is possible to set a stimulation of 0.1 ms, 0.3 ms or 1.0 ms.

The selected setting is displayed and stored. Setting can be changed by pressing the "ms" button when the LED display is lit.

Image 7
Contents Stimuplex HNS Manufacturer Distributor Stockert GmbH Contents Checks before start-up Page Technical description Selecting the pulse width Switching on / switching offOperating the nerve stimulator Selecting the pulse frequencySelecting the current display mode Checking the battery voltageSelecting the current range Pulse width Current display modeCurrent adjustment range LED displaysPulse LED display LCD displayTesting the Stimuplex HNS 11 before an application Initial set up, maintenance, and general adviceTone Technical data Special technical featuresChanging the battery BatteryMaintenance and safety checks Cleaning and disinfecting the Stimuplex HNSPage Warranty Error messages Safety informationStimuplex D needles 30 bevel Instruments and accessoriesProduct discription Canula Code number Sales unit Length Stimuplex D needles, 15 bevelContiplex D needles, 30 bevel Product discription Canula Code numberLength Stimuplex a needles, 30 bevel Contiplex D needles, 15 bevelPrinciples Peripheral electrical nerve stimulationPage Page Literature Appendix B Appendix aAppendix C Only 0.2 or 0.05 mA Stimulation current required Appendix DPage Hospital Care B. Braun Melsungen AG