12
Special technical features
The configuration of the electrode connection provides protection against a wrong polarity connection of the electrode cable.
This new type of plug connection system ensures correct polarity of the electrodes at all times.
A high stimulation voltage is associated with an unusually wide range of adjustment with extremely small stimulation electrodes.
The Stimuplex® HNS 11 generates a negative,
Use of the latest microcomputer technology has made it possible to integrate performance characteristics which until now could only be provided in large systems. For example: selection of cur- rent display mode, set and actual value display, display of battery charge, error messages.
An internal quartz time base ensures precise pulse widths and frequencies.
Technical data
Type: Stimuplex® HNS 11 nerve stimulator
Instrument type | BF |
Battery | 9 V |
Power consumption | 3.3 mA |
Stimulation current | max. 5 mApp / 0 - 12 k |
Stimulation voltage | max. 65 Vpp |
Stimulation frequency .... | 1 Hz / 2 Hz |
Measuring tolerance | Adjustment control display =3% (set point) |
| Flowing current display = 2% (actual value) |
| based on set mA max values (5 mA or 1 mA) |
Weight | 250 g |
Due to its high maximum output voltage the nerve stimulator cov- ers a wide range of skin and needle impedance. This is especially important when using stimulation needles with an extremely small conductive area (pin point electrode tip) such as Stimuplex® D and Contiplex® D needles.