Teledyne 3000TA operating instructions #### % Span Slope=#### ppm/s

Page 60

Operation

3000TA- EU

 

 

 

Use the ▲/▼ keys to toggle between AUTO and MAN span settling. Stop when MAN appears, blinking, on the display. Press Enter to move to the next screen.

Press Enter to move to the next screen.

Calib. Holding time

Cal hold: 5 min

This menu allows the operator to set the time the analyzer should be held in the auto span mode. It does not affect anything in Manual Mode. Just press Enter to continue.

Span Val: 000008.00 <ENT>Span <UP>Mod #

Press ▲ (<UP>) to permit modification (Mod #) of span value.

Use the arrow keys to enter the oxygen concentration of the span gas you are using (209000.00 if you are using air). The ◄►arrows choose the digit, and the ▲/▼ arrows choose the value of the digit.

Press Enter to enter the span value into the system and begin the span calibration.

Once the span has begun, the microprocessor samples the output at a predetermined rate. It calculates the difference between successive samplings and displays this difference as Slope on the screen. It takes several seconds for the first Slope value to display. Slope indicates rate of change of the Span reading. It is a sensitive indicator of stability.

#### % Span

Slope=#### ppm/s

When the Span value displayed on the screen is sufficiently stable, press Enter. (Generally, when the Span reading changes by 1% or less of the full scale of the range being calibrated for a period of ten minutes it is sufficiently stable.) Once Enter is pressed, the Span reading changes to the correct value. The instrument then automatically enters the Analyze function.

Teledyne Analytical Instruments

48

Image 60
Contents Model 3000TA Teledyne Analytical Instruments3000TA- EU Copyright 2011 Teledyne Analytical InstrumentsSpecific Model Information Trace Oxygen AnalyzerImportant Notice Safety Messages 3000TA- EU Vii Table of Contents Maintenance OperationAppendix List of Figures Model 3000TA Front PanelList of Tables Main Features of the Analyzer IntroductionOverview Typical ApplicationsModel Designations Introduction 3000TA- EUFront Panel Operator Interface Model 3000TA Front PanelRecognizing Difference Between LCD & VFD Rear Panel Equipment Interface ∙ 50-Pin Equipment Interface PortIntroduction 3000TA- EU Operational Theory IntroductionMicro-Fuel Cell Sensor Trace Oxygen Analyzer Operational TheoryOperational Theory 3000TA- EU Anatomy of a Micro-Fuel CellElectrochemical Reactions Calibration Characteristics Effect of PressureCharacteristic Input/Output Curve for a Micro-Fuel Cell Sample SystemPiping Layout and Flow Diagram for Standard Model Electronics and Signal Processing Flow Diagram3000TA Internal Electronic Component Location Block Diagram of the Model 3000TA-EU Electronics Operational Theory 3000TA- EU Installation Trace Oxygen Analyzer InstallationUnpacking the Analyzer Mounting the AnalyzerInstallation 3000TA- EU Front Panel of the Model 3000TARear Panel Connections Gas ConnectionsSample Exhaust OUT Electrical ConnectionsEquipment Interface Connector Pin Arrangement ∙ Threshold Alarm Alarm Relay Contact Pins Pin Function Range ID Relay Connections Remote Probe Connections 2.3 RS-232 PortCommands via RS-232 Input Installing the Micro-Fuel Cell Testing the SystemPage Using the Data Entry and Function Buttons OperationTrace Oxygen Analyzer Operation Operation 3000TA- EU System Function Hierarchy of Functions and Sub functionsTracking Oxygen Readings During Calibration and Alarm Delay TRAK/HLD Auto-Cal Pswd Logout More Setting up an Auto-Cal Password Protection A a a Characters Available for Password Definition Logout System Self-Diagnostic Test Version Screen Showing Negative Oxygen ReadingsZero and Span Functions Zero Cal Zero Settling Man ENT To Begin Span Cal Span Val 000008.00 ENTSpan UPMod # #### % Span Slope=#### ppm/s Alarms Function Span FailureAL-1AL-2 Choose Alarm Range Function Setting the Analog Output Ranges Fixed Range Analysis Signal Output Analyze FunctionMED Operation 3000TA- EU Teledyne Analytical Instruments Maintenance Routine MaintenanceTrace Oxygen Analyzer Maintenance Storing and Handling Replacement CellsMaintenance 3000TA- EU When to Replace a CellRemoving the Micro-Fuel Cell Installing a New Micro-Fuel Cell Removing the Micro-FuelMaintenance 3000TA- EU Fuse Replacement Cell WarrantySystem Self Diagnostic Test Installing FusesMajor Internal Components Cleaning TroubleshootingTrace Oxygen Analyzer Maintenance Maintenance 3000TA- EU Teledyne Analytical Instruments Model 3000TA Specifications AppendixTrace Oxygen Analyzer Appendix Recommended 2-Year Spare Parts List Appendix 3000TA- EUOrders should be sent to Drawing List 19-inch Relay Rack Panel MountApplication notes Flow Rate Recommendations3000TA Examples Bypass Conversions Material Safety Data Sheet Section I Product IdentificationSection II Physical and Chemical Data Appendix 3000TA- EU Section III -Physical Hazards Section IV Health Hazard DataSection VI Handling Information
Related manuals
Manual 19 pages 17.56 Kb

3000TA specifications

The Teledyne 3000TA is an advanced telemetry system designed for high-performance applications in various industries such as aerospace, automotive, and research. This innovative device combines cutting-edge technologies with user-friendly features, making it an essential tool for engineers and scientists alike.

One of the standout features of the Teledyne 3000TA is its robust data acquisition capabilities. It offers a high sampling rate, which allows users to capture rapid changes in measurements accurately. This is particularly beneficial for applications that require real-time data analysis and decision-making, such as testing the performance of new aircraft components or monitoring engine behavior under different loads.

The 3000TA integrates multiple sensor interfaces, enabling it to connect seamlessly with various types of sensors, including temperature, pressure, and strain gauges. This versatility makes it suitable for a wide range of testing scenarios, from structural analysis to environmental monitoring. Users can customize the system to fit their specific needs, significantly enhancing its functionality.

Another significant advantage of the Teledyne 3000TA is its advanced wireless communication technology. By employing protocols like Bluetooth and Wi-Fi, it allows for remote data monitoring and access, which improves safety and convenience during testing processes. Engineers can easily gather data from hazardous environments without being physically present, ensuring both efficiency and safety.

The device is also equipped with an intuitive user interface, which simplifies the setup and operation processes. The touchscreen display provides real-time data visualizations and easy navigation through various menus, making it accessible even to those new to telemetry systems. This user-centric design reduces the learning curve and enhances overall productivity.

In terms of data integrity, the Teledyne 3000TA features robust error-checking and validation protocols. These built-in safeguards ensure that the data collected is accurate and reliable, which is crucial for making informed decisions based on telemetry data.

The system's compact and lightweight design ensures portability, allowing for use in various fields and environments. This flexibility enhances its appeal for field applications, where space and weight can be significant constraints.

In summary, the Teledyne 3000TA stands out as a versatile and reliable telemetry solution, combining high-performance data acquisition, advanced communication technologies, and user-friendly features. Its application across multiple industries makes it an invaluable tool for professionals seeking to optimize their testing and analysis processes.