Teledyne Maintenance 3000TA- EU, When to Replace a Cell, Removing the Micro-Fuel Cell

Page 70

Maintenance

3000TA- EU

 

 

 

around the unit. What may appear to be plain water could contain one of these toxic substances. In case of eye contact, immediately flush eyes with water for at least 15 minutes. Call physician. (See appendix, Material Safety Data Sheet.)

CAUTION: DO NOT DISTURB THE INTEGRITY OF THE CELL PACKAGE UNTIL THE CELL IS TO ACTUALLY BE USED. IF THE CELL PACKAGE IS PUNCTURED AND AIR IS PERMITTED TO ENTER, THE CELL WILL REQUIRE AN EXCESSIVELY LONG TIME TO REACH ZERO AFTER INSTALLATION (1-2 WEEKS!).

5.2.2 When to Replace a Cell

The characteristics of the Micro-Fuel Cell show an almost constant output throughout its useful life and then fall off sharply towards zero at the end. Cell failure in the 3000TA is usually characterized inability to zero the instrument down to a satisfactorily low ppm reading. When this occurs, the 3000TA system alarm trips, and the LCD displays a failure message.

#.# ppm Anlz

CELL FAIL/ ZERO HIGH

Before replacing the cell:

a)Check your span gas to make sure it is within specifications.

b)Check for leaks downstream from the cell, where oxygen may be leaking into the system.

If there are no leaks and the span gas is OK, replace the cell.

5.2.3 Removing the Micro-Fuel Cell

The Micro-Fuel cell is located inside the stainless steel cell block behind the front panel (see Figure 5-1). To remove an existing cell:

1.Remove power to the instrument by unplugging the power cord at the power source.

2.Open the front panel door by pressing the release button on the top right corner of the door all the way in with a narrow gauge tool, such as a small screwdriver, and releasing it.

Teledyne Analytical Instruments

58

Image 70
Contents Model 3000TA Teledyne Analytical Instruments3000TA- EU Copyright 2011 Teledyne Analytical InstrumentsSpecific Model Information Trace Oxygen AnalyzerImportant Notice Safety Messages 3000TA- EU Vii Table of Contents Maintenance OperationAppendix List of Figures Model 3000TA Front PanelList of Tables Overview Main Features of the AnalyzerIntroduction Typical ApplicationsModel Designations Introduction 3000TA- EUFront Panel Operator Interface Model 3000TA Front PanelRecognizing Difference Between LCD & VFD Rear Panel Equipment Interface ∙ 50-Pin Equipment Interface PortIntroduction 3000TA- EU Micro-Fuel Cell Sensor Operational TheoryIntroduction Trace Oxygen Analyzer Operational TheoryOperational Theory 3000TA- EU Anatomy of a Micro-Fuel CellElectrochemical Reactions Calibration Characteristics Effect of PressureCharacteristic Input/Output Curve for a Micro-Fuel Cell Sample SystemPiping Layout and Flow Diagram for Standard Model Electronics and Signal Processing Flow Diagram3000TA Internal Electronic Component Location Block Diagram of the Model 3000TA-EU Electronics Operational Theory 3000TA- EU Unpacking the Analyzer InstallationTrace Oxygen Analyzer Installation Mounting the AnalyzerInstallation 3000TA- EU Front Panel of the Model 3000TARear Panel Connections Gas ConnectionsSample Exhaust OUT Electrical ConnectionsEquipment Interface Connector Pin Arrangement ∙ Threshold Alarm Alarm Relay Contact Pins Pin Function Range ID Relay Connections Remote Probe Connections 2.3 RS-232 PortCommands via RS-232 Input Installing the Micro-Fuel Cell Testing the SystemPage Operation Using the Data Entry and Function ButtonsTrace Oxygen Analyzer Operation Operation 3000TA- EU System Function Hierarchy of Functions and Sub functionsTracking Oxygen Readings During Calibration and Alarm Delay TRAK/HLD Auto-Cal Pswd Logout More Setting up an Auto-Cal Password Protection A a a Characters Available for Password Definition Logout System Self-Diagnostic Test Version Screen Showing Negative Oxygen ReadingsZero and Span Functions Zero Cal Zero Settling Man ENT To Begin Span Cal Span Val 000008.00 ENTSpan UPMod # #### % Span Slope=#### ppm/s Alarms Function Span FailureAL-1AL-2 Choose Alarm Range Function Setting the Analog Output Ranges Fixed Range Analysis Signal Output Analyze FunctionMED Operation 3000TA- EU Teledyne Analytical Instruments Trace Oxygen Analyzer Maintenance MaintenanceRoutine Maintenance Storing and Handling Replacement CellsWhen to Replace a Cell Maintenance 3000TA- EURemoving the Micro-Fuel Cell Installing a New Micro-Fuel Cell Removing the Micro-FuelMaintenance 3000TA- EU Fuse Replacement Cell WarrantySystem Self Diagnostic Test Installing FusesMajor Internal Components Cleaning TroubleshootingTrace Oxygen Analyzer Maintenance Maintenance 3000TA- EU Teledyne Analytical Instruments Appendix Model 3000TA SpecificationsTrace Oxygen Analyzer Appendix Recommended 2-Year Spare Parts List Appendix 3000TA- EUOrders should be sent to Drawing List 19-inch Relay Rack Panel MountApplication notes Flow Rate Recommendations3000TA Examples Bypass Conversions Section I Product Identification Material Safety Data SheetSection II Physical and Chemical Data Appendix 3000TA- EU Section III -Physical Hazards Section IV Health Hazard DataSection VI Handling Information
Related manuals
Manual 19 pages 17.56 Kb

3000TA specifications

The Teledyne 3000TA is an advanced telemetry system designed for high-performance applications in various industries such as aerospace, automotive, and research. This innovative device combines cutting-edge technologies with user-friendly features, making it an essential tool for engineers and scientists alike.

One of the standout features of the Teledyne 3000TA is its robust data acquisition capabilities. It offers a high sampling rate, which allows users to capture rapid changes in measurements accurately. This is particularly beneficial for applications that require real-time data analysis and decision-making, such as testing the performance of new aircraft components or monitoring engine behavior under different loads.

The 3000TA integrates multiple sensor interfaces, enabling it to connect seamlessly with various types of sensors, including temperature, pressure, and strain gauges. This versatility makes it suitable for a wide range of testing scenarios, from structural analysis to environmental monitoring. Users can customize the system to fit their specific needs, significantly enhancing its functionality.

Another significant advantage of the Teledyne 3000TA is its advanced wireless communication technology. By employing protocols like Bluetooth and Wi-Fi, it allows for remote data monitoring and access, which improves safety and convenience during testing processes. Engineers can easily gather data from hazardous environments without being physically present, ensuring both efficiency and safety.

The device is also equipped with an intuitive user interface, which simplifies the setup and operation processes. The touchscreen display provides real-time data visualizations and easy navigation through various menus, making it accessible even to those new to telemetry systems. This user-centric design reduces the learning curve and enhances overall productivity.

In terms of data integrity, the Teledyne 3000TA features robust error-checking and validation protocols. These built-in safeguards ensure that the data collected is accurate and reliable, which is crucial for making informed decisions based on telemetry data.

The system's compact and lightweight design ensures portability, allowing for use in various fields and environments. This flexibility enhances its appeal for field applications, where space and weight can be significant constraints.

In summary, the Teledyne 3000TA stands out as a versatile and reliable telemetry solution, combining high-performance data acquisition, advanced communication technologies, and user-friendly features. Its application across multiple industries makes it an invaluable tool for professionals seeking to optimize their testing and analysis processes.