Teledyne 3000TA-XL-EU quick start Calibration of the Analyzer, Auto Mode Zero Calibration

Page 13

Model 3000 TA-XL-EU

Quickstart Guide

 

 

◄► arrow keys to scroll back and forth between letters, and the ▲▼ arrow keys to change the letters to the proper password.

Press Enter to enter the password.

Change Password? <ENT>=Yes <ESC>=No

Press Escape.

Calibration of the Analyzer

For trace analysis, the analyzer must be purged before calibration. Refer to the Instruction manual for purging procedure.

Zero Calibration

Make sure an appropriate zero gas is connected to the sample port for this procedure.

Auto Mode Zero Calibration

Press Zero to enter the zero function mode.

Use the ▲▼ arrow keys to toggle between AUTO and MAN zero settling. Stop when AUTO appears blinking.

Zero: Settling: AUTO

<ENT> To Begin

Press Enter to begin zeroing.

####PPM Zero

Slope=#### ppm/s

.

.

.

####PPM Zero

4 Left=### ppm/s

Teledyne Analytical Instruments

11

Image 13
Contents Model 3000TA-XL-EU Page Front Panel Operator Interface Getting StartedFunction Keys Access DoorData Entry Keys Digital Meter DisplayGas Inlet and Outlet One inlet and one exhaust out Rear Panel Equipment InterfaceMake Gas Connections Mount the AnalyzerInstall the fuse Connect primary input powerElectrical Connections Install 50-Pin Equipment Interface Connector Install the Micro-fuel CellConnect the 9-pin D connector to the RS-232 serial port 2CXL Micro-fuel CellModel 3000 TA-XL-EU Quickstart Guide Using the Data Entry and Function Buttons Default ConfigurationAuto Ranging on Available Functions Entering the Password Password ProtectionCalibration of the Analyzer Auto Mode Zero CalibrationZero Calibration Auto Mode Spanning Manual Mode ZeroingSpan Calibration Manual Mode Spanning Setting Alarm ParametersSet The Range Function Fixed Range or Autoranging Analysis Analyze Function