Teledyne 3000TA-XL-EU quick start Setting Alarm Parameters, Manual Mode Spanning

Page 15

Model 3000 TA-XL-EU

Quickstart Guide

 

 

####ppm Span

Slope=#### ppm/s

The instrument automatically returns to the Analyze mode at the conclusion of the auto span cycle..

Manual Mode Spanning:

Press Span to start the Span function.

Span: Settling:MAN

<ENT> For Next

Use the ▲▼ keys to toggle between AUTO and MAN span settling. Stop when MAN appears, blinking, on the display. Press Enter to move to the next screen.

Calib. Holding time

Cal hold: 5 min

Press Enter key to continue.

Span Val: 000008.00 <ENT>Span <UP>Mod #

Press ▲ (<UP>) to permit modification (Mod #) of span value.

Use the arrow keys to enter the oxygen concentration of the span gas you are using (209000.00 if you are using air). The ◄► arrows choose the digit, and the ▲▼ arrows choose the value of the digit.

Press Enter to enter the span value into the system and begin the span calibration.

####% Span

Slope=#### ppm/s

When the Span value displayed on the screen is sufficiently stable, press Enter and the span reading changes to the correct value.

The instrument automatically returns to the Analyze mode.

Setting Alarm Parameters

If you are using password protection, you will need to enter your password to access the alarm functions. Follow the instructions above for Entering the Password to enter your password. Once you have clearance to proceed, you can set the alarms.

Teledyne Analytical Instruments

13

Image 15
Contents Model 3000TA-XL-EU Page Front Panel Operator Interface Getting StartedDigital Meter Display Access DoorFunction Keys Data Entry KeysGas Inlet and Outlet One inlet and one exhaust out Rear Panel Equipment InterfaceMake Gas Connections Mount the AnalyzerConnect primary input power Install the fuseElectrical Connections 2CXL Micro-fuel Cell Install the Micro-fuel CellInstall 50-Pin Equipment Interface Connector Connect the 9-pin D connector to the RS-232 serial portModel 3000 TA-XL-EU Quickstart Guide Default Configuration Using the Data Entry and Function ButtonsAuto Ranging on Available Functions Entering the Password Password ProtectionAuto Mode Zero Calibration Calibration of the AnalyzerZero Calibration Manual Mode Zeroing Auto Mode SpanningSpan Calibration Manual Mode Spanning Setting Alarm ParametersSet The Range Function Fixed Range or Autoranging Analysis Analyze Function