Model 3000 | Quickstart Guide |
|
|
####ppm Span
Slope=#### ppm/s
The instrument automatically returns to the Analyze mode at the conclusion of the auto span cycle..
Manual Mode Spanning:
•Press Span to start the Span function.
Span: Settling:MAN
<ENT> For Next
•Use the ▲▼ keys to toggle between AUTO and MAN span settling. Stop when MAN appears, blinking, on the display. Press Enter to move to the next screen.
Calib. Holding time
Cal hold: 5 min
•Press Enter key to continue.
Span Val: 000008.00 <ENT>Span <UP>Mod #
•Press ▲ (<UP>) to permit modification (Mod #) of span value.
•Use the arrow keys to enter the oxygen concentration of the span gas you are using (209000.00 if you are using air). The ◄► arrows choose the digit, and the ▲▼ arrows choose the value of the digit.
•Press Enter to enter the span value into the system and begin the span calibration.
####% Span
Slope=#### ppm/s
•When the Span value displayed on the screen is sufficiently stable, press Enter and the span reading changes to the correct value.
The instrument automatically returns to the Analyze mode.
Setting Alarm Parameters
If you are using password protection, you will need to enter your password to access the alarm functions. Follow the instructions above for Entering the Password to enter your password. Once you have clearance to proceed, you can set the alarms.
Teledyne Analytical Instruments | 13 |