Teledyne 3000TA-XL-EU quick start Set The Range Function

Page 16

Quickstart Guide

Trace Oxygen Analyzer

 

 

Press the Alarm button on the front panel to enter the Alarm function. Make sure that AL–1is blinking.

AL—1 AL—2

Choose Alarm

Using the ◄► arrow keys to move the blinking over to AL–1. Then press Enter.

AL—1 1000 ppm HI

Dft—N Fs—N Ltch—N

Five parameters can be changed on this screen:

Value of the alarm setpoint, AL–1 #### ppm (oxygen); value can be set from 0.00 to 10,000.00 ppm

Out-of-range direction, HI or LO

Defeated? Dft–Y/N(Yes/No)

Failsafe? Fs–Y/N(Yes/No)

Latching? Ltch–Y/N(Yes/No).

To change any of the alarm parameters, use the ◄► arrow keys to move the blinking over to the specific parameter.

Use the ▲▼ arrow keys to change the value of the parameter. Holding down the key speeds up the incrementing or decrementing.

Once the parameters for alarm 1 have been set, press Alarms again, and repeat this procedure for alarm 2 (AL–2).

To reset a latched alarm, go to Dft– and then press either ▲ two times or ▼ two times. (Toggle it to Y and then back to N.)

—or—

Go to Ltch– and then press either ▲ two times or ▼ two times. (Toggle it to N and back to Y.)

Set The Range Function

To set the 3 programmable ranges:

Enter the range function mode by pressing the Range button on the front panel.

L—1 M—10

H—100 Mode—AUTO

Teledyne Analytical Instruments

14

Image 16
Contents Model 3000TA-XL-EU Page Getting Started Front Panel Operator InterfaceAccess Door Function KeysData Entry Keys Digital Meter DisplayRear Panel Equipment Interface Gas Inlet and Outlet One inlet and one exhaust outMount the Analyzer Make Gas ConnectionsInstall the fuse Connect primary input powerElectrical Connections Install the Micro-fuel Cell Install 50-Pin Equipment Interface ConnectorConnect the 9-pin D connector to the RS-232 serial port 2CXL Micro-fuel CellModel 3000 TA-XL-EU Quickstart Guide Using the Data Entry and Function Buttons Default ConfigurationAuto Ranging on Available Functions Password Protection Entering the PasswordCalibration of the Analyzer Auto Mode Zero CalibrationZero Calibration Auto Mode Spanning Manual Mode ZeroingSpan Calibration Setting Alarm Parameters Manual Mode SpanningSet The Range Function Analyze Function Fixed Range or Autoranging Analysis