Teledyne 3000TA-XL-EU quick start Manual Mode Zeroing, Span Calibration, Auto Mode Spanning

Page 14

Quickstart Guide

Trace Oxygen Analyzer

 

 

The zeroing process will automatically conclude when the output is within the acceptable range for a good zero. Then the analyzer automatically returns to the Analyze mode.

Manual Mode Zeroing

Press Zero to enter the Zero function.

Use the ▲▼ keys to toggle between AUTO and MAN zero settling. Stop when MAN appears, blinking, on the display.

Zero: Settling: Man

<ENT> To Begin

Press Enter to begin the zero calibration.

####ppm Zero

Slope=#### ppm/s

Once zero settling is completed, the instrument automatically returns to the Analyze mode.

Span Calibration

Make sure an appropriate span gas is connected to the sample port.

Auto Mode Spanning:

Press Span to enter the span function.

Use the arrow keys to toggle between AUTO and MAN span settling. Stop when AUTO appears, blinking on the display.

Span: Settling: AUTO

<ENT> For Next

Press Enter.

Calib. Holding time

Cal hold: 5 min

Press Enter to move to the next screen.

Span Val: 000008.00 <ENT>Span <UP>Mod #

Use the ▲▼ arrow keys to enter the oxygen-concentration mode. The ◄► arrows choose the digit, and the ▲▼ arrows choose the value of the digit. When you have finished typing in the concentration of the span gas you are using, press Enter to begin the span calibration.

Teledyne Analytical Instruments

12

Image 14
Contents Model 3000TA-XL-EU Page Getting Started Front Panel Operator InterfaceData Entry Keys Access DoorFunction Keys Digital Meter DisplayRear Panel Equipment Interface Gas Inlet and Outlet One inlet and one exhaust outMount the Analyzer Make Gas ConnectionsElectrical Connections Connect primary input powerInstall the fuse Connect the 9-pin D connector to the RS-232 serial port Install the Micro-fuel CellInstall 50-Pin Equipment Interface Connector 2CXL Micro-fuel CellModel 3000 TA-XL-EU Quickstart Guide Auto Ranging on Default ConfigurationUsing the Data Entry and Function Buttons Available Functions Password Protection Entering the PasswordZero Calibration Auto Mode Zero CalibrationCalibration of the Analyzer Span Calibration Manual Mode ZeroingAuto Mode Spanning Setting Alarm Parameters Manual Mode SpanningSet The Range Function Analyze Function Fixed Range or Autoranging Analysis