Precision Trading PM5300, PM5200 user manual Specifications

Page 6

Specifications

Model

 

PM5200 High Flow

PM5300 Low Flow

 

 

 

 

 

 

 

Primary Outlet

 

 

15 - 120 lpm

3 - 30 lpm

Flow Range

 

With both supply pressures at 50 psi (3.4 bar)

 

 

 

 

with NO BLEED

 

Auxiliary Outlet

 

 

2 - 100 lpm

0 - 30 lpm

Flow Range

 

 

 

 

 

 

 

 

With both supply pressures at 50 psi (3.4 bar)

 

 

 

 

 

 

with BLEED

 

Bleed Flow

 

13 lpm or less

3 lpm or less

 

 

at

50 psi (3.4 bar)

at 50 psi (3.4 bar)

Maximum

 

 

 

 

 

 

 

Combined

 

 

120 lpm

30 lpm

Flow (All Outlets)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Bypass Flow

 

 

 

 

 

 

 

(Loss of Air or

 

 

> 85 lpm

> 45 lpm

Oxygen supply)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Bypass Alarm

 

50 psi

60 psi

50 psi

 

60 psi

Activation

 

(3.45 bar)

(4.14 bar)

(3.45 bar)

 

(4.14 bar)

 

 

13-25 psi

16-24 psi

18-22 psi

 

16-24 psi

 

 

 

 

 

 

 

 

0.9-1.7 bar

1.1-1.65 bar

1.2-1.5 bar

1.1-1.65 bar

 

 

 

 

 

 

 

 

 

 

 

 

Alarm Reset:

 

When pressure differential is

 

 

 

 

6 psi (0.4 bar) or less.

 

Alarm Sound Level:

 

to 80 db at 1 ft (0.3 m)

 

Oxygen Concentration

21 - 100%

 

 

 

Adjustment Range:

 

 

 

 

 

 

Gas Supply Pressure:

30 - 75 psi (2.1 - 5.2 bar)

 

 

 

 

Air and Oxygen within 10 psi (0.67 bar)

 

 

 

of each other

 

 

 

Mixed Gas Stability:

±1% Oxygen

 

 

 

 

 

 

Connection Types:

 

DISS Type - Air & Oxygen Inlets & Outlets

 

 

 

and / or

 

 

 

 

 

 

NIST Type - Air & Oxygen Inlets

Blender

Air-Oxygen

Image 6
Contents E R M a N U a L Contents Intended Use Read ALL Instructions Before UsingReceiving / Inspection Explanation of AbbreviationsSafety Information Warnings and Cautions Consult Accompanying DocumentsBlender Specifications Transport / Storage Requirements Pressure DropDryness and Composition for inlet gases Diagrams Component Description Air Inlet Fitting InstallationAlarm Rear Slide Mount with dove tailAlarm Test Reverse Gas Flow ProcedureOperating Instructions Cleaning Performance CheckVerification consists Maintenance Technical Description Returns Disposal InstructionsPlease Recycle Troubleshooting Problem Probable Cause RemedyLimited Warranty Limitation of Liability Two 2 years from shipmentDeclaration of Conformity