Specifications continued
Dimensions: (without fittings)
Depth: | 4.9 in | (12.5 cm) |
Width: | 2.3 in | (5.7 cm) |
Height: | 4.1 in | (10.4 cm) |
Weight: | 2.9 lbs | (1.3 kg) |
Shipping Weight: | 3.5 lbs | (1.6 kg) |
Operating Temperature Range: | 59°F to 104°F (15°C to 40°C) |
Transport / Storage Requirements
Temperature Range: | |
Humidity: | Max 95% Noncondensing |
FiO2 Accuracy:* | ± 3% of full scale @ 50 psi (3.4 bar) |
| ± 4% of full scale @ 60 psi (4.14 bar) |
Pressure Drop:
Low Flow: ≤ 2 psi (0.14 bar) at inlet pressures from
High Flow: ≤ 3 psi (0.21 bar) at inlet pressures from
The
The
The Oxygen Analyzer should comply with ISO 7767 to meet CE requirement.
Dryness and Composition for inlet gases:
Air: Medical Air supply should meet the requirements of ANSI Z86.1 - 1973 commodity specification for Air, type 1 grade D or better.
Oxygen: Oxygen supply must meet all requirements of USP Medical Oxygen Grade N.
Dew Point:
(ONLY for CE requirements)
Both inlets should remain 10°F (5.5°C) or more below the lowest temperature to which the air distribution system equipment is exposed. At a temperature of 25°F
*Accuracy of oxygen concentration will be affected if bleed is not activated at flow settings below 15 lpm for the high flow Blender, and 3 lpm for the low flow Blender.
Specifications are subject to change without prior notice.