3.To activate the bleed, turn the knurled collar until the bleed pin is engaged. Then slide the collar back until it contacts the cover.
4.Confirm the concentration of air/oxygen with an Oxygen Analyzer/Monitor.
5.Confirm the flow of air and/or oxygen mixture to the patient.
6.Turn “OFF” the Air and Oxygen supply when deactivating the Auxiliary bleed, and no device is attached to the outlet port.
7.To deactivate the bleed, pull the collar out away from the cover while rotating the collar until the slot engages the pin. Then slide the collar past the pin and turn an additional 1/4 turn to lock in place.
NOTE: This Procedure is completed with ease when all gas supplies to the Blender are turned “OFF”.
8.Turn “OFF” the Air and Oxygen supply or disconnect when the Blender is not in use.
9.Operational Verification listed below should be done before the Blender is placed in service.
Verification consists of:
•Alarm Test
•Verify FiO2 Concentration with a Oxygen Analyzer/Monitor
•Reverse Gas Flow Procedure
Performance Check
A detailed description of the tests listed below can be found in the Blender Service Manual (P/N 504827), available on the Internet; www.precisionmedical.com
•Reverse Gas Flow Procedure
•Operational Verification Procedure (OVP)
CLEANING
CAUTION
•DO NOT steam autoclave.
•DO NOT immerse the
•DO NOT use any strong solvent or abrasive cleaners.
•DO NOT gas sterilize with (ETO) Ethylene Trioxide.
•DO NOT clean with aromatic hydrocarbons.
Blender |
|
| |
| |
11 | |
|