Precision Trading PM5300, PM5200 user manual Performance Check, Cleaning, Verification consists

Page 12

3.To activate the bleed, turn the knurled collar until the bleed pin is engaged. Then slide the collar back until it contacts the cover.

4.Confirm the concentration of air/oxygen with an Oxygen Analyzer/Monitor.

5.Confirm the flow of air and/or oxygen mixture to the patient.

6.Turn “OFF” the Air and Oxygen supply when deactivating the Auxiliary bleed, and no device is attached to the outlet port.

7.To deactivate the bleed, pull the collar out away from the cover while rotating the collar until the slot engages the pin. Then slide the collar past the pin and turn an additional 1/4 turn to lock in place.

NOTE: This Procedure is completed with ease when all gas supplies to the Blender are turned “OFF”.

8.Turn “OFF” the Air and Oxygen supply or disconnect when the Blender is not in use.

9.Operational Verification listed below should be done before the Blender is placed in service.

Verification consists of:

Alarm Test

Verify FiO2 Concentration with a Oxygen Analyzer/Monitor

Reverse Gas Flow Procedure

Performance Check

A detailed description of the tests listed below can be found in the Blender Service Manual (P/N 504827), available on the Internet; www.precisionmedical.com

Reverse Gas Flow Procedure

Operational Verification Procedure (OVP)

CLEANING

CAUTION

DO NOT steam autoclave.

DO NOT immerse the Air-Oxygen Blender into any liquid.

DO NOT use any strong solvent or abrasive cleaners.

DO NOT gas sterilize with (ETO) Ethylene Trioxide.

DO NOT clean with aromatic hydrocarbons.

Blender

 

 

 

Air-Oxygen

11

 

Image 12
Contents E R M a N U a L Contents Read ALL Instructions Before Using Receiving / InspectionIntended Use Explanation of AbbreviationsSafety Information Warnings and Cautions Consult Accompanying DocumentsBlender Specifications Transport / Storage Requirements Pressure DropDryness and Composition for inlet gases Diagrams Component Description Installation AlarmAir Inlet Fitting Rear Slide Mount with dove tailAlarm Test Reverse Gas Flow ProcedureOperating Instructions Cleaning Performance CheckVerification consists Maintenance Technical Description Returns Disposal InstructionsPlease Recycle Troubleshooting Problem Probable Cause RemedyLimited Warranty Limitation of Liability Two 2 years from shipmentDeclaration of Conformity