WARNING
•An Oxygen Analyzer/Monitor must be used to verify oxygen concentration.
•Accuracy of oxygen concentration will be affected if bleed is not activated at flow settings below 15 lpm for the High Flow Blender, and 3 lpm for the Low Flow Blender.
•DO NOT obstruct the alarm.
•DO NOT use Blender when alarm is sounding.
•DO NOT use oil in or around the Blender.
•DO NOT occlude or obstruct the bleed port on the auxiliary outlet of the Blender.
•DO NOT use near any type of flame or flammable/ explosive substances, vapors or atmosphere.
•Oxygen Concentration Dial does not rotate 360 degrees. Rotating the dial less than 21% or over 100% oxygen will damage the Blender.
CAUTION
•Turn off gas supplies when
•Store the
•The
•Ensure all connections are tight and leak free.
•Avoid excessive pressure surges greater than 100 psi (6.9 bar) when pressuring the Blender inlets.
•DO NOT steam autoclave.
•DO NOT immerse
•DO NOT gas sterilize with (ETO) Ethylene Trioxide.
•DO NOT use if dirt or contaminants are present on or around the Blender or connecting devices.
•DO NOT smoke in an area where oxygen is being administered.
•DO NOT clean with aromatic hydrocarbons.