Invacare IRC5LXO2, IRC10LXO2 manual Specifications

Page 13

SECTION 4—SPECIFICATIONS

SECTION 4—SPECIFICATIONS

Electrical Requirements:

120 VAC ± 10% (132 VAC/108 VAC), 60

 

Hz

 

 

Rated Current Input:

Platinum 5 and XL: 4.3 A

 

 

 

Platinum 10: 4.7 A

 

 

Sound Level:

Platinum 5 and XL: 50 dB Average

 

 

 

Platinum 10: 58 dB Average

 

 

Altitude:

Platinum 5 and XL - Up to 6,000 ft. (1828

 

meters) above sea level without degradation

 

of concentration levels. From 6,000 ft. (1828

 

meters) to 13,129 ft. (4000 meters) below

 

90% efficiency

 

 

 

Platinum 10 - Up to 4,000 ft. (1230 meters)

 

above sea level without degradation of con-

 

centration levels. From 4,000 ft. (1230

 

meters) to 13,129 ft. (4000 meters) below

 

90% efficiency

 

 

*Oxygen Output

All 5LXO2/5LX models 95.6% to 87% at 0.5

Concentration Levels:

to 5 L/min. and all 10LXO2/10LX models

 

95.6% to 87% at 2 L to 10 L/min.

 

 

Maximum Outlet Pressure:

Platinum 5 and XL: 5 psi ± 0.5 psi. (34.5 kPa

 

± 3.45 kPa.)

 

 

 

Platinum 10: 9 psi ± 0.5 psi. (62.0 kPa ± 3.45

 

kPa.)

 

 

*NOTE: Stated concentration levels achieved after initial warm-up period (approximately 30 minutes).

Part No 1118353

13

PlatinumSeries

Image 13
Contents PlatinumSeries XL, 5 Accessories Table of Contents Maintenance Special Notes General Guidelines General Guidelines Radio Frequency Interference MaintenanceImmediately Polarized Plug Instruction Accessories not Shown FeaturesUnpacking HandlingInspection StorageSpecifications Label Symbol O2 Purity Indicator Lights Specifications Flow rates PlatinumSeries PlatinumSeries Please Seal with Tape Before Mailing Product Registration Form Product was purchased for use at Home Facility Page Introduction OperatingSelect a Location Recommended Guidelines for Optimum Performance Plug in Power CordHumidifier Compartment Connect Humidifier If So PrescribedAttaching the Humidifier Bottle Adapter Power Switch Flowrate Power SwitchInitial Startup of the Concentrator Explanation of Oxygen Purity Indicator LightsLX models indicator Light Explanation Indicators If Your Unit Does Not Feature the O2 SensorNormal I/O Routine Maintenance MaintenanceCabinet Preventive Maintenance Record Solution Troubleshooting GuideTroubleshooting Guide Symptom LOW Flow Alarm Limited Warranty Limited Warranty USA
Related manuals
Manual 32 pages 52.1 Kb