Invacare IRC5LXO2, IRC10LXO2 manual LOW Flow Alarm

Page 37

SECTION 7—TROUBLESHOOTING GUIDE

SYMPTOM:

PROBABLE CAUSE:

SOLUTION:

 

 

 

Alarm:

LOW FLOW ALARM

1a. Inspect for kinks or

RAPID

1a. Kinked or blocked

blockages. Correct, clean or

 

tubing, cannula or humid-

replace item. Once

Beep....Beep...

ifier.

corrected, turn power OFF

Beep...Beep

 

for 60 seconds and then turn

 

1b. Flowmeter set at 0.5

power back ON.

 

L/min. for Platinum 5 and

 

 

1 L/min. for Platinum 10

1b. Reset flowmeter to pre-

 

or less.

scribed flowrate. If prescribed

 

 

flowrate is below 0.5 L/min.,

 

 

use the Pediatric Flowmeter

 

 

IRCPF16 (Only with Platinum

 

 

5 and XL).

 

 

 

Part No 1118353

37

PlatinumSeries

Image 37
Contents PlatinumSeries XL, 5 Accessories Table of Contents Maintenance Special Notes General Guidelines General Guidelines Radio Frequency Interference MaintenanceImmediately Polarized Plug Instruction Accessories not Shown FeaturesUnpacking HandlingInspection StorageSpecifications Label Symbol O2 Purity Indicator Lights Specifications Flow rates PlatinumSeries PlatinumSeries Please Seal with Tape Before Mailing Product Registration Form Product was purchased for use at Home Facility Page Introduction OperatingSelect a Location Recommended Guidelines for Optimum Performance Plug in Power CordHumidifier Compartment Connect Humidifier If So PrescribedAttaching the Humidifier Bottle Adapter Power Switch Flowrate Power SwitchInitial Startup of the Concentrator Explanation of Oxygen Purity Indicator LightsLX models indicator Light Explanation Indicators If Your Unit Does Not Feature the O2 SensorNormal I/O Routine Maintenance MaintenanceCabinet Preventive Maintenance Record Solution Troubleshooting GuideTroubleshooting Guide Symptom LOW Flow Alarm Limited Warranty Limited Warranty USA
Related manuals
Manual 32 pages 52.1 Kb