Invacare IRC5LXO2, IRC10LXO2 manual Flowrate, Power Switch

Page 28

SECTION 5—OPERATING INSTRUCTIONS

On/Off (I/O)

Power Switch

FIGURE 5.5 Power Switch

Flowrate

NOTE: For this procedure, refer to FIGURE 5.6 on page 29.

1.Turn the flowrate knob to the setting prescribed by your physician or therapist.

WARNING

DO NOT change the L/min. setting on the flowmeter unless a change has been prescribed by your physician or therapist.

NOTE: To properly read the flowmeter, locate the prescribed flowrate line on the flowmeter. Next, turn the flow knob until the ball rises to the line. Now, center the ball on the L/min. line prescribed.

2.If the flowrate on the flowmeter ever falls below 0.5 L/min. for Platinum 5 (1 L/min. for Platinum 10) for more than about one minute, the LOW FLOW alarm will be triggered. This is a rapid beeping of the audible alarm. Check your tubing or accessories for blocked or kinked tubing or a defective humidifier bottle. After rated flow is restored, more than 0.5 L/min. for Platinum 5 (1 L/min. for Platinum 10), the LOW FLOW audible alarm will go off.

NOTE: The use of some accessories such as the pediatric flowstand and the HomeFill II compressor will deactivate the Low Flow Alarm.

PlatinumSeries

28

Part No 1118353

Image 28
Contents PlatinumSeries XL, 5 Accessories Table of Contents Maintenance Special Notes General Guidelines General Guidelines Maintenance Radio Frequency InterferenceImmediately Polarized Plug Instruction Features Accessories not ShownHandling UnpackingInspection StorageSpecifications Label Symbol O2 Purity Indicator Lights Specifications Flow rates PlatinumSeries PlatinumSeries Please Seal with Tape Before Mailing Product Registration Form Product was purchased for use at Home Facility Page Introduction OperatingSelect a Location Plug in Power Cord Recommended Guidelines for Optimum PerformanceConnect Humidifier If So Prescribed Humidifier CompartmentAttaching the Humidifier Bottle Adapter Power Switch Power Switch FlowrateExplanation of Oxygen Purity Indicator Lights Initial Startup of the ConcentratorIndicators If Your Unit Does Not Feature the O2 Sensor LX models indicator Light ExplanationNormal I/O Maintenance Routine MaintenanceCabinet Preventive Maintenance Record Troubleshooting Guide SolutionTroubleshooting Guide Symptom LOW Flow Alarm Limited Warranty Limited Warranty USA
Related manuals
Manual 32 pages 52.1 Kb