SECTION 6—MAINTENANCE
SECTION 6—MAINTENANCE
NOTE: The Invacare concentrators are specifically designed to minimize routine preventive maintenance. In places with high dust or soot levels, maintenance may need to be performed more often. Only authorized or
NOTE: At a minimum, preventive maintenance MUST be performed according to the maintenance record guidelines. Refer to Preventive Maintenance Record on page 34.
Routine Maintenance
NOTE: The following routine maintenance should be performed more frequently. See recommended intervals in each section and preventive maintenance record.
Cleaning the Cabinet Filter
NOTE: For this procedure, refer to FIGURE 6.1.
NOTE: There are two cabinet filters one located on each side of the cabinet.
1.Remove each filter and clean at least once a week depending on environmental conditions.
NOTE: Environmental conditions that may require more frequent
cleaning of the filters include but are not limited to: high dust, smoking, air pollutants, etc.
2.Clean the cabinet filters with a vacuum cleaner or wash in warm soapy water and rinse thoroughly.
3.Dry the filters thoroughly before reinstallation.
Platinum™Series | 32 | Part No 1118353 |