SECTION 1—GENERAL GUIDELINES

WARNING

The use of oxygen therapy requires that special care be taken to reduce the risk of fire. Any materials that will burn in air, and some that will not, are easily ignited and burn rapidly in high concentrations of oxygen. For safety concerns, it is necessary that all sources of ignition be kept away from the product and preferably out of the room in which it is being used. NO SMOKING signs should be prominently displayed.

A spontaneous and violent ignition may occur if oil, grease or greasy substances come in contact with oxygen under pressure. These substances MUST be kept away from the oxygen concentrator, tubing and connections, and all other oxygen equipment. DO NOT use any lubricants unless recommended by Invacare.

For optimum performance, Invacare recommends that each concentrator be on and running for a minimum of 30 minutes at a time. Shorter periods of operation may reduce maximum product life.

If the concentrator has a damaged cord or plug, if it is not working properly, if it has been dropped or damaged, or dropped into water, call qualified technician for examination and repair.

Keep the oxygen tubing, cord, and unit away from heated or hot surfaces, including space heaters, blankets, stoves and similar electrical appliances.

DO NOT move or relocate concentrator by pulling on the cord.

NEVER drop or insert any object into any opening.

Part No 1118353

7

PlatinumSeries

Image 7
Contents PlatinumSeries XL, 5 Accessories Table of Contents Maintenance Special Notes General Guidelines General Guidelines Radio Frequency Interference MaintenanceImmediately Polarized Plug Instruction Accessories not Shown FeaturesStorage HandlingUnpacking InspectionSpecifications Label Symbol O2 Purity Indicator Lights Specifications Flow rates PlatinumSeries PlatinumSeries Please Seal with Tape Before Mailing Product Registration Form Product was purchased for use at Home Facility Page Introduction OperatingSelect a Location Recommended Guidelines for Optimum Performance Plug in Power CordHumidifier Compartment Connect Humidifier If So PrescribedAttaching the Humidifier Bottle Adapter Power Switch Flowrate Power SwitchInitial Startup of the Concentrator Explanation of Oxygen Purity Indicator LightsLX models indicator Light Explanation Indicators If Your Unit Does Not Feature the O2 SensorNormal I/O Routine Maintenance MaintenanceCabinet Preventive Maintenance Record Solution Troubleshooting GuideTroubleshooting Guide Symptom LOW Flow Alarm Limited Warranty Limited Warranty USA