ResMed VPAP IV ST manual Technical Specifications, Symbols Which Appear On The Device

Page 18

Technical Specifications

Operating pressure range

2 to 25 cm H2O

Maximum single fault steady

40 cm H2O

state pressure

 

Pressure measurement

±0.5 cm H2O ± 4% of the measured reading

tolerance

 

Flow measurement tolerance

±0.1 or 20% of reading, whichever is greater

S, ST and T modes

IPAP: 4 to 25 cm H2O (measured at the mask); EPAP: 2 to 25 cm H2O

 

(measured at the mask); Pressure Support: 0 to 23 cm H2O

CPAP mode

4 to 20 cm H2O (measured at the mask)

DECLARED DUAL-NUMBER NOISE EMISSION VALUES in accordance with ISO 4871:

Sound pressure level

26 dBA with uncertainty of 2 dBA as measured according to

 

ISO 17510-1: 2002

 

28 dBA with uncertainty of 2 dBA as measured according to

 

ISO 17510-1: 2007

Sound power level

36 dBA with uncertainty of 2 dBA as measured according to

 

ISO 17510-1: 2007

Dimensions (L x W x H)

112 mm x 164 mm x 145 mm

Weight

1.3 kg

Power supply

Input range 100–240V, 50–60Hz, 40VA (typical power consumption),

 

< 100VA (maximum power consumption)

Housing construction

Flame retardant engineering thermoplastic

Operating temperature

+5ºC to +35ºC

Operating humidity

10–95% non-condensing

Storage and transport

-20ºC to +60ºC

temperature

 

Storage and transport humidity

10–95% non-condensing

Operating altitude

Sea level to 2,600 m

Electromagnetic compatibility

Product complies with all applicable electromagnetic compatibility

 

requirements (EMC) according to IEC60601-1-2, for residential,

 

commercial, and light industry environments

Air filter

Two-layered, powder-bonded, polyester non-woven fiber

Air tubing

Flexible plastic, 1 x 2 m

Air outlet

The 22 mm conical air outlet complies with ISO 5356-1

IEC 60601-1 classification

Class II (double insulation), Type CF

Note:

The manufacturer reserves the right to change these specifications without notice.

Pressure may be displayed in cm H20 or hPa.

Symbols Which Appear On The Device

Caution;

Follow instructions for use;

Drip proof;

Type CF equipment;

Dangerous voltage; Class II equipment; Start/Stop and mask-fit; Manufacturer.

English

Technical Specifications

15

Image 18
Contents User Guide Vpap Vpap IV ST Contents Contraindications IntroductionVpap System Adverse EffectsResScan Data Card MasksHumidifier Supplemental OxygenUsing the Vpap IV and Vpap IV ST Setting up your VpapHow to Use the Control Panel Attaching a H4i HumidifierUsing the Menus ServicingHow to Select the Mask Type SmartStartStarting Treatment Using the Mask-Fitting feature Stopping TreatmentReminders on the Vpap LCD Smart DataUsing the Data Card Copying Data onto a Data CardTraveling with the Vpap IV and Vpap IV ST Weekly Cleaning and MaintenanceDaily MonthlyServicing Troubleshooting Problem/Possible Cause SolutionComponent failure Technical Specifications Symbols Which Appear On The DeviceGeneral Warnings and Cautions General Warnings and Cautions Page Limited Warranty