Calculation Method for Displayed Values

<Pulse level meter>

The pulse level meter displays the transmitted light amount ratio calculated by the following equation. The pulse level is indicated in eight levels (0 to 15%, full scale if over 10%).

Pulse level (%) =

Transmitted light amount that varies due to the pulse

x 100

------------------------------------------------------------------------------------------------------------------Transmitted light amount that does not vary due to the pulse

<Oxygen saturation (SpO2)>

Updates the moving average of the oxygen saturation over the last 12 seconds at one-second intervals.

<Pulse rate (P.R.)>

Updates the moving average of the last 8 pulse rates at one-second intervals.

Appendix

25

Image 27
Contents PULSOX-300 Safety Symbols Safety Precautions Foreword Contents Foreword Safety NotesAbout Probe Package ContentsSystem Diagrams and Accessories Names of Parts Foreword Difference from the SaO2 value to displayed value Foreword Disposal Method Preparations PreparationsConnecting the probe Installing the battery Starting the Measurement MeasurementPulse rate P.R Reading the displayEnding the Measurement About Auto Power-Off functionError Messages Trouble ShootingsCheck Points Check Points Check Points Check Points before repairing Display SpecificationProbes SR-5C, SP-5C, LM-5C, SD-5C Calculation Method for Displayed Values AppendixEMC Guidance Compliance Measurement Principle Oxygen Saturation vs. Oxygen Partial Pressure Manufacturer