Invacare 211EX, 20EX, 212EX, 50EX manual If using an Oxygen Cylinder and Regulator

Page 17

OPERATION

3.If using an oxygen cylinder, check that the regulator is properly attached before opening the cylinder valve.

4.Connect the Venture DODD to the regulator by performing one (1) of the following:

A.If using an Oxygen Cylinder and Regulator:

On a 20 psig regulator, connect the tubing (supplied with the Venture DODD) from the barb fitting of the regulator to the barb inlet fitting on the Venture DODD.

On a 50 psig regulator, connect the threaded (DISS) fitting (supplied with the Venture DODD) to the regulator DISS fitting.

NOTE: The 50 psig system should ONLY be used with the threaded (DISS) fitting and not the 12-inch connecting tube, as the pressure may be too great for the tube to stay connected.

B.If using a Liquid Oxygen Reservoir:

Directly connect the Venture DODD to the oxygen source using a DISS fitting.

5.Connect the nasal cannula tubing to the barbed oxygen outlet fitting on the Venture DODD and adjust cannula for comfort.

WARNING

When using nasal or masked type administering equipment, oxygen or air tubing MUST be routed and secured properly to ensure that the tubing does NOT become entangled, kinked or severed.

NOTE: Use ONLY a NASAL CANNULA with PULSE MODE

delivery, NOT a mask. A mask may NOT fit your face well enough for the unit to sense your inhalation.

NOTE: A special nasal cannula is NOT required. Any standard flow crush-proof nasal cannula can be used.

NOTE: A MASK or a CANNULA may be used with CONTINUOUS FLOW.

NOTE: The cannula tubing can be up to 50 feet long to function properly in the pulse mode. The continuous flow mode is NOT affected by the length of the canula tubing.

PART NO. 1093845

17

VENTURE™

Image 17
Contents Demand Oxygen Delivery Device and System Venture Table of Contents Safety Summary To Prevent Fire Risk of FireOther Precautions Radio Frequency Interference Unpacking / Handling UnpackingInspection Handling ProceduresIntroduction Invacare Venture Demand Oxygen Delivery DeviceStatus Indicators FeaturesVenture Dodd Features Venture Dodd System Features Alarms SpecificationsEnvironmental Operation USE of OxygenSET-UP and Normal Pulse Operation If using an Oxygen Cylinder and Regulator Switching to Continuous Flow Operation Maintenance Prolonging the Life of Your BatteryCleaning the Outer Case Accessories and Spare Parts Problem Possible Cause Solutions TroubleshootingCall your Home Health External input will not Incorrect power adapter Oxygen Conservation Oxygen Conserved and Battery USE Time Based on Breath RateLimited Warranty Canada