TROUBLESHOOTING
|
| PROBLEM | POSSIBLE CAUSE | SOLUTIONS | ||
|
|
|
|
|
|
|
O2 LED - (GREEN) | Indicates Oxygen Pulse Flow | This is normal | ||||
Blinking. | in normal pulse delivery. | operation. | ||||
|
|
|
|
|
|
|
O2 LED - (GREEN) | Indicates Venture DODD | Turn unit OFF, check | ||||
|
|
|
|
| not sensing presence of | cylinder pressure is above |
|
|
|
|
| oxygen in pulse delivery. | refill limit (500 psig) and |
|
|
|
|
|
| post valve is open, check |
|
|
|
|
|
| regulator is set to correct |
|
|
|
|
|
| flowrate. If problem still |
|
|
|
|
|
| exists, call your Home |
|
|
|
|
|
| Health Care Dealer or |
|
|
|
|
|
| Service Representative |
|
|
|
|
|
| IMMEDIATELY and go to |
|
|
|
|
|
| continuous mode if |
|
|
|
|
|
| oxygen supply is adequate. |
|
|
|
|
|
|
|
O2 and | LED - | Indicates oxygen pulse | Breathe normally. Check | |||
(GREEN and YELLOW) | delivery, but Venture DODD | cannula for kinking and | ||||
Blinking |
| not sensing inhalation. | check tubing connection. | |||
|
|
|
|
|
|
|
| LED - (YELLOW) | Indicates no pulse | Breathe normally. Check | |||
Blinking and | delivery, Venture DODD | cannula for kinking and | ||||
Intermittent Beep | not sensing inhalation. | check tubing connection. | ||||
|
|
|
|
|
|
|
| LED - (YELLOW) | Possible internal problem. | Turn unit off then on | |||
Solid and Continuous | End user possibly not | again. If problem still | ||||
Beep |
| receiving oxygen. | exists, call your Home | |||
|
|
|
|
|
| Health Care Dealer or |
|
|
|
|
|
| Service Representative |
|
|
|
|
|
| IMMEDIATELY and go to |
|
|
|
|
|
| continuous mode if |
|
|
|
|
|
| oxygen supply is adequate. |
|
|
|
|
|
|
|
|
|
|
| LED - (RED) | Battery low. | Replace battery as soon |
|
|
|
| |||
|
|
|
| |||
Blink | ing | , Three (3) Beeps | Approximately two (2) hours | as possible. | ||
at Beginning. | of battery use time left. |
| ||||
|
|
|
|
|
|
|
|
|
|
| LED - (RED) Solid, | Approximately 30 minutes | Replace battery or turn |
|
|
|
| |||
|
|
|
| |||
Six ( | 6) | Beeps at Beginning. | of battery use time left. | unit OFF to enable | ||
|
|
|
|
|
| continuous mode if |
|
|
|
|
|
| oxygen supply is adequate. |
|
|
|
|
|
|
|
PART NO. 1093845 | 23 | VENTURE™ |