Invacare 212EX Oxygen Conserved and Battery USE Time Based on Breath Rate, Oxygen Conservation

Page 26

OXYGEN CONSERVATION

OXYGEN CONSERVED AND BATTERY USE TIME BASED ON BREATH RATE

 

BREATH RATE (BPM)

% OXYGEN CONSERVED

10

20

30

40

 

 

 

 

 

Variable Breath Mode:

82

77

70

63

Fixed 1.0 sec. Breath Mode:

83

66

50

33

Fixed 0.75 sec. Breath Mode:

87

75

62

50

Fixed 0.50 sec. Breath Mode:

91

83

75

66

 

 

 

 

 

i.e. 77% Oxygen conservation over Continuous Flow achieved with Venture DODD in Variable Breath Mode @ 20 BPM Rate. Oxygen supply lasts 4.3-times longer. Venture DODD responds to individual breath rates. Therefore, use times stated are only approximates. Actual savings depends on individual breathing patterns.

COMPATIBILITY

Venture DODD requires using a flowrate controlled oxygen source. Gaseous or liquid oxygen systems that meet the following requirements are compatible to the Venture DODD:

Venture 20 requires a 20 +3 psig (138 + 20 kPa) flow source. Venture 50 requires a 50 +5 psig (345 + 35 kPa) flow source. Flowrate Range: 0-6 L/Min.

VENTURE™

26

PART NO. 1093845

Image 26
Contents Demand Oxygen Delivery Device and System Venture Table of Contents Safety Summary Risk of Fire To Prevent FireOther Precautions Radio Frequency Interference Unpacking Unpacking / HandlingHandling Procedures InspectionInvacare Venture Demand Oxygen Delivery Device IntroductionFeatures Status IndicatorsVenture Dodd Features Venture Dodd System Features Specifications AlarmsEnvironmental Operation USE of OxygenSET-UP and Normal Pulse Operation If using an Oxygen Cylinder and Regulator Switching to Continuous Flow Operation Prolonging the Life of Your Battery MaintenanceCleaning the Outer Case Accessories and Spare Parts Troubleshooting Problem Possible Cause SolutionsCall your Home Health External input will not Incorrect power adapter Oxygen Conserved and Battery USE Time Based on Breath Rate Oxygen ConservationLimited Warranty Canada