OPERATION
6.Set the prescribed flow rate on the oxygen source by performing one (1) of the following:
A.If using an Oxygen Cylinder and Regulator:
OPEN cylinder valve and adjust regulator to prescribed flow rate.
B.If using a Liquid Oxygen Reservoir:
Set flow control knob on reservoir to prescribed flowrate.
7.Press the ON button to engage pulse mode. The unit will pulse each time the user inhales, the GREEN PULSE LED will blink with each oxygen pulse delivery.
8.As battery power depletes, the unit will alarm to inform the end user to replace the battery or switch to external power source.
A.INITIAL BATT. LED - The Red LED will FLASH and the audible alarm will BEEP three (3) times. This indicates that you have about two (2) hours of use time left on your battery.
B.SECOND BATT. LED - The Red LED will be SOLID and the audible alarm will BEEP six (6) times. This indicates that you have between 15 and 30 minutes of use time left on your battery.
NOTE: Remaining battery use times are dependent on battery type being used. Alkaline battery type is recommended for longest life.
OFF
(Continuous Mode
Operation)
FIGURE 3 - SWITCHING TO CONTINUOUS FLOW
VENTURE™ | 18 | PART NO. 1093845 |