SAFETY SUMMARY

WARNING (CONTINUED)

DO NOT wear the Demand Oxygen Delivery Device under any clothing.

OTHER PRECAUTIONS:

Never attempt to lubricate the oxygen delivery system.

DO NOT insert any object (except suitable AC/DC Adapter) in external power jack of Demand Oxygen Delivery Device. This may cause an electrical shock which could cause burns.

DO NOT subject the Demand Oxygen Delivery Device to temperatures below -40°F (-40°C) or above 158°F (70°C). To do so could damage the electronics and battery.

NEVER tamper with the oxygen delivery system. Doing so could create a hazardous condition and possibly cause the oxygen delivery system to malfunction. If you have any questions, call your Home Health Care Dealer or Service Representative.

ALWAYS connect the Demand Oxygen Delivery Device to a pressure regulated oxygen source. ONLY connect 20 psig Demand Oxygen Delivery Device to a 20 psig oxygen source and a 50 psig Demand Oxygen Delivery Device to a 50 psig oxygen source.

DO NOT use any additional lengths of tubing between the oxygen source and the Demand Oxygen Delivery Device. ONLY use Invacare tubing supplied with the Demand Oxygen Delivery Device, if any tubing is required.

DO NOT use a mask when using pulse mode delivery, USE ONLY a standard nasal cannula.

When using the continuous flow delivery, a mask or nasal cannula can be used.

VENTURE™

6

PART NO. 1093845

Image 6
Contents Demand Oxygen Delivery Device and System Venture Table of Contents Safety Summary Risk of Fire To Prevent FireOther Precautions Radio Frequency Interference Unpacking Unpacking / HandlingHandling Procedures InspectionInvacare Venture Demand Oxygen Delivery Device IntroductionFeatures Status IndicatorsVenture Dodd Features Venture Dodd System Features Specifications AlarmsEnvironmental USE of Oxygen SET-UP and Normal Pulse OperationOperation If using an Oxygen Cylinder and Regulator Switching to Continuous Flow Operation Prolonging the Life of Your Battery MaintenanceCleaning the Outer Case Accessories and Spare Parts Troubleshooting Problem Possible Cause SolutionsCall your Home Health External input will not Incorrect power adapter Oxygen Conserved and Battery USE Time Based on Breath Rate Oxygen ConservationLimited Warranty Canada