SAFETY SUMMARY

WARNING (CONTINUED) the supplier and/or physician should be contacted IMMEDIATELY. No adjustments should be made to the flowrate unless prescribed by a physician.

DO NOT attempt to alter the switches in the battery compartment. These switches are preset by your Home Health Care Dealer and control the breath delivery prescription.

RISK OF FIRE:

Oxygen does not burn. However, it will accelerate burning of flammable materials.

TO PREVENT FIRE:

DO NOT SMOKE when operating the oxygen delivery system. No Smoking signs should be prominently displayed.

Keep oxygen delivery system at least 5-feet away from ANY electrical appliances; such as radios, TVs, air conditioners, fans, electric razors, etc.

Keep the oxygen delivery system away from open flames or heating sources such as furnaces or stoves.

Keep flammable materials away from the oxygen delivery system; such as oils, grease and aerosol sprays. These will ignite and burn quickly in the presence of concentrated oxygen.

TO PREVENT HIGH CONCENTRATIONS OF OXYGEN: Keep the oxygen delivery system in a well ventilated area at all times.

When the oxygen delivery system is not in use, press the OFF button to place the Demand Oxygen Delivery Device in the Continuous Mode, turn the Cylinder OFF, bleed the remaining oxygen from the Regulator, turn the Regulator OFF.

PART NO. 1093845

5

VENTURE™

Image 5
Contents Demand Oxygen Delivery Device and System Venture Table of Contents Safety Summary To Prevent Fire Risk of FireOther Precautions Radio Frequency Interference Unpacking / Handling UnpackingInspection Handling ProceduresIntroduction Invacare Venture Demand Oxygen Delivery DeviceStatus Indicators FeaturesVenture Dodd Features Venture Dodd System Features Alarms SpecificationsEnvironmental Operation USE of OxygenSET-UP and Normal Pulse Operation If using an Oxygen Cylinder and Regulator Switching to Continuous Flow Operation Maintenance Prolonging the Life of Your BatteryCleaning the Outer Case Accessories and Spare Parts Problem Possible Cause Solutions TroubleshootingCall your Home Health External input will not Incorrect power adapter Oxygen Conservation Oxygen Conserved and Battery USE Time Based on Breath RateLimited Warranty Canada