Instruction Manual
| Model NGA2000 TO2 | |
| March 2003 | |
| 
 | Network Connections | 
| 
 | Power Connection | 
| 
 | Fuse | 
| 
 | Power Indicator Light | 
Figure 2-4.  Trace Oxygen Analyzer Front Panel
2-5  ELECTRICAL CONNECTIONS
NOTE
Electrical connnections must be made in compliance with National Electrical Code (ANSI/NFPA 70) and/or any applicable na- tional or electical codes.
Two electrical connections are required on the Analyzer Module: POWER and NETWORK (See Figure 
either of which is appropriate for: 1) intercon- nection with the Backplane of the Platform or 2) 
Connect a source of 24 V 5A DC power to the power inlet. Make sure that the ground connec- tion is made, and that this is separate from the power return lead. Failure to ensure a good ground may result in random noise and distur- bance in the analyzer readings.
| Rosemount Analytical Inc. A Division of Emerson Process Management | Installation |