Emerson NGA2000 TO2 instruction manual GAS Overpressure, Sample Flow

Page 20

Instruction Manual

748374-F March 2003

Model NGA2000 TO2

Sample

Inlet

1/4" Tube

Sample

Exhaust

1/4" Tube

Figure 2-3. Trace Oxygen Analyzer Back Panel Connections

CAUTION

GAS OVERPRESSURE

At no time should sample, zero or span gas inlet pressure exceed 69 hPa - gauge (1.0 psig). Damage to the Sensor may occur if this pressure level is exceeded.

CAUTION

SAMPLE FLOW

Do not test the sample pressure by blocking the exhaust. When the pres- sure is released the sudden surge of flow will spin the internal flowmeter off its bearings and destroy it.

c.LEAK TEST

The TO2 Analyzer Module is completely tested at the factory for gas leakage. The user is responsible for testing for leakage only at the inlet and outlet fittings on the rear panel. Caution: Do not expose the

Sensor to pressure in excess of 1.0 psig as this may cause damage.

2-4

Installation

Rosemount Analytical Inc. A Division of Emerson Process Management

Image 20
Contents Model NGA2000 TO2 Essential Instructions Table of Contents Model NGA2000 TO2Material Safety Data Sheet Maintenance and TroubleshootingList of Illustrations List of TablesModel NGA2000 TO2 Preface Intended USE Statement DefinitionsSafety Summary Authorized PersonnelModel NGA2000 TO2 Model NGA2000 TO2 Documentation CompliancesGlossary of Terms Section Description and Specifications FeaturesOverview Typical ApplicationsTrace Oxygen Detector Technology Trace Oxygen Analyzer Module Top View Specifications GeneralSample PhysicalSection Installation AssemblyUnpacking LocationConnections PressureCalibration Gases Trace Oxygen Analyzer Outline and Mounting Dimensions GAS Overpressure Sample FlowTrace Oxygen Analyzer Front Panel Electrical ConnectionsModel NGA2000 TO2 Displays Run Mode DisplaySection Startup and Operation Basic Controls Menu Startup Procedure Quick Start FeatureGAS Scale Factor GSF CalibrationRoutine Operation Description Type Alarm IndicationTrace Oxygen Analyzer Module Alarms Model NGA2000 TO2 Refilling Sensor Section Maintenance and TroubleshootingOverview Qualified Personnel Electrolyte Replacement Sensor ReplacementWater Addition To add water FusesFlow Sensor Replacement TroubleshootingPrinted Circuit Boards Sensor DamageTO2 analyzer gives erratic and very in- sensitive readings Section Replacement Parts Configuration IdentifierReplacement Parts MatrixModel NGA2000 TO2 Customer Service Rosemount Analytical Inc Customer Service CenterSection Return of Material Return of MaterialModel NGA2000 TO2 NGA TO2 Analyzer Module Menus Section Appendix A. Menu StructureNGA TO2 Menu Structure Introduction Model NGA2000 TO2 Model NGA2000 TO2 Model NGA2000 TO2 Model NGA2000 TO2 Model NGA2000 TO2 Path Section Appendix B. User Interface HelpMenu Items Model NGA2000 TO2 Section Index Model NGA2000 TO2 Warranty Emerson Process Management Rosemount Analytical Inc