Emerson NGA2000 TO2 instruction manual Section Startup and Operation, Displays, Run Mode Display

Page 23

Model NGA2000 TO2

Instruction Manual

748374-F March 2003

SECTION 3

STARTUP AND OPERATION

3-1 OVERVIEW

Prior to initial startup, the user should perform the leak test procedure outlined in Section 2.

For the remainder of this section, Analyzer Module interconnection with a Platform or some interfacing component will be assumed. Display and Keypad information refers to that which the user can expect to see and do with regard to the Front Panel of the Platform.

(For a complete description of Platform Front Panel controls and indicators, see the Platform instruction manual.

3-2 DISPLAYS

Three kinds of Display screens are available to the user:

Run Mode

Menu

Help

a.Run Mode Display

The Run Mode is the normal mode of op- eration. In this mode, the Display will show the current gas measurement, the compo- nent of interest, the current operations of the softkeys, and several graphics: a bar representing the displayed concentration as a percent of fullscale and up to four lines showing user selectable secondary parame- ters from either the Analyzer Module or any IO module bound to it. See the Platform manual for information as to how to select these.

If more than one Analyzer Module is con- nected to the system, an additional Run Mode display will show as many as four (five for version 2.3 and later) gas meas- urements on screen.

TO2 1

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

19.4

 

 

 

 

 

 

 

ppm O2

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

0 Range 3

 

 

 

 

 

 

 

 

 

 

50

Sensor current:

1.4403 mA

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Sensor temperature:

 

 

 

26.3 C

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Temperature current:

2.01 mA

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Sample Flow:

511 ml/min

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

DISPLAY

 

PARMS.

 

 

 

MENU

 

NEXT

 

 

INFO

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Figure 3-1. Run Mode Display

b.Menu Displays

The Menu structure enables the user to ac- cess data and functions, and put informa- tion onto the network. From the Run Mode display, press the MENUS softkey to gain access to the Main Menu.

Main Menu

Basic controls . . .

Expert controls and set up . . .

(Operational configuration)

Technical level configuration . . .

(Diagnostic and manufacturing/service)

DISPLAY

 

PARMS.

 

NEXT

 

EXT LOCK

 

INFO

 

 

 

 

 

 

 

 

 

Figure 3-2. Main Menu

The Main Menu is subdivided into three lev- els of control based generally on which per- sonnel is likely to use it: Basic Controls - Operators, Expert Controls and set up - System Engineers, and Technical level con- figuration - Analyzer technicians. Many lay- ers of the menu structure are described at appropriate places throughout this manual.

Rosemount Analytical Inc. A Division of Emerson Process Management

Startup and Operation

3-1

Image 23
Contents Model NGA2000 TO2 Essential Instructions Model NGA2000 TO2 Table of ContentsMaintenance and Troubleshooting Material Safety Data SheetList of Tables List of IllustrationsModel NGA2000 TO2 Definitions Preface Intended USE StatementAuthorized Personnel Safety SummaryModel NGA2000 TO2 Model NGA2000 TO2 Compliances DocumentationGlossary of Terms Typical Applications Section Description and SpecificationsFeatures OverviewTrace Oxygen Detector Technology Trace Oxygen Analyzer Module Top View Physical SpecificationsGeneral SampleLocation Section InstallationAssembly UnpackingConnections PressureCalibration Gases Trace Oxygen Analyzer Outline and Mounting Dimensions Sample Flow GAS OverpressureElectrical Connections Trace Oxygen Analyzer Front PanelModel NGA2000 TO2 Displays Run Mode DisplaySection Startup and Operation Basic Controls Menu Calibration Startup ProcedureQuick Start Feature GAS Scale Factor GSFRoutine Operation Description Type Alarm IndicationTrace Oxygen Analyzer Module Alarms Model NGA2000 TO2 Refilling Sensor Section Maintenance and TroubleshootingOverview Qualified Personnel Fuses Electrolyte ReplacementSensor Replacement Water Addition To add waterSensor Damage Flow Sensor ReplacementTroubleshooting Printed Circuit BoardsTO2 analyzer gives erratic and very in- sensitive readings Matrix Section Replacement PartsConfiguration Identifier Replacement PartsModel NGA2000 TO2 Return of Material Customer ServiceRosemount Analytical Inc Customer Service Center Section Return of MaterialModel NGA2000 TO2 NGA TO2 Analyzer Module Menus Section Appendix A. Menu StructureNGA TO2 Menu Structure Introduction Model NGA2000 TO2 Model NGA2000 TO2 Model NGA2000 TO2 Model NGA2000 TO2 Model NGA2000 TO2 Path Section Appendix B. User Interface HelpMenu Items Model NGA2000 TO2 Section Index Model NGA2000 TO2 Warranty Emerson Process Management Rosemount Analytical Inc