
Instruction Manual
Model NGA2000 TO2
| with a clean, metallic diaphragm,  | Pressure | |
| regulator. A shutoff valve is recommended. | Constant between 13.8 and 69 hPa - gauge | |
| 
 | ||
| Calibration Gases | (0.2 and 1.0 psig) sample inlet pressure is | |
| recommended. If a needle valve is used | ||
| 
 | ||
| The TO2 module does not require routine | upstream of the Analyzer Module to control | |
| zero calibration. The zero is factory set and | flow, the inlet pressure to the needle valve | |
| does not experience routine drift. Over long | should not exceed 345 hPa (5 psig). A | |
| periods of time, the zero may experience | constant sample flow rate between 1.0 to | |
| minor drift. For low ppm range analyzers, | 3.0 SCFH (0.5 to 1.5 l/min) is recommended | |
| you may wish to check the zero at one year | for best results. The Analyzer Module must | |
| intervals.  | vent to atmosphere to avoid back pressure | |
| mended for use as zero gas. This gas is | influences on the oxygen reading. | |
| certified to <0.5 ppm oxygen and can be | 
 | |
| improved by passing the zero gas through | b. Connections | |
| an oxygen scrubber such as Millipore™ | 
 | |
| Waferpure or Semigas Nanochem® resin | (See Figure  | |
| purifiers. A mixture of trace oxygen in a | and outlet lines for sample to appropriately | |
| background of nitrogen is recommended as | labeled fittings on the rear panel. SAMPLE | |
| span gas. For maximum accuracy, the con- | IN and SAMPLE OUT are  | |
| centration of trace oxygen in the span gas | type compression fittings. Zero and span | |
| should be as high as possible for the range | gases should be introduced at the SAMPLE | |
| of measurement. | IN fitting at normal sample inlet flow rate. | |
| Sample | Metallic tubing is recommended for the | |
| The sample must be clean and dry before | sample line. The use of plastic, Teflon, or | |
| other  | ||
| entering the Analyzer Module. Sample | ||
| bient oxygen permeation through the tubing | ||
| should be filtered for particulates down to | ||
| causing higher than expected reading. Ex- | ||
| two microns, and should have a dewpoint at | ||
| haust tubing should be 1/4 inch (6.3 mm) or | ||
| least 5 °C (13 °F) below the coldest ex- | ||
| larger, and can be metallic or  | ||
| pected ambient temperature. | ||
| 
 | 
| ANALYZER MODULE | BACKPLANE | 
| CONNECTIONS | CONNECTIONS | 
| Network 1 | Backplane | 
| 
 | |
| Network 2 | 
 | 
| Power | Controller Board | 
| 
 | Connector | 
| Fuse | Network | 
| Power Indicator Light | |
| 
 | Power | 
Figure 2-1.  Analyzer Module Interconnection with Instrument Platform
| Installation | Rosemount Analytical Inc. A Division of Emerson Process Management |