Evolution Technologies OM-900M manual Description of Parts & Controls

Page 11

DESCRIPTION OF PARTS & CONTROLS

Low Battery Indicator Light: The EVOLUTIONTM MOTION conserver incorporates a visual indicator light that alerts you when a battery change is needed.. The following signal will be displayed as your batteries become depleted:

Blinking red light - Low battery energy level.. Replace batteries..

Cylinder Adjustment Handle: This is used to attach the unit to any CGA 870 post-valve cylinder..

Selector Knob: This enables the user to select their prescribed oxygen flow setting.. It also enables s/he to select the continuous flow mode setting.. The continuous flow mode setting (CF) is designed for emergency use only.. The amount of oxygen delivered when using the EVOLUTIONTM MOTION in continuous flow mode is preset at 2 LPM (liters per minute).. When not in use,the cylinder should be turned off by turning the cylinder valve toggle counter clockwise..

The CF mode (continuous flow) is a mechanical override designed as a safety feature in case battery power fails.. It is set at

2 LPM CF.. When in the CF mode, the user can still change the LED display lights by pressing the Selector Button but it has no effect on oxygen output..

CAUTION! In the event that it is necessary to operate the unit in the continuous flow mode, DO NOT obstruct the flow of oxygen from the Oxygen Supply Outlet by placing your finger over the outlet or blocking the flow through the oxygen tubing in any way.. Doing so may render the unit inoperable and/or damage the sensor in the unit..

*NOTE: Remember that in continuous flow mode, the oxygen will be consumed at a much faster rate.

Return to another source before depleting the oxygen cylinder.

Battery Holder: This compartment holds two (2) AA-size alkaline batteries..

Oxygen Pressure Gauge: This enables the user to monitor the contents of the compressed oxygen cylinder..

Oxygen Supply Outlet: Use this fitting to attach a standard cannula..

Cylinder Alignment Pins: When assembling the unit, these parallel pins must go into the holes on the post valve..

Seal Washer: This creates the interface between the valve and the EVOLUTIONTM MOTION conserver.. Besides offering a rugged interface, it also surrounds the oxygen path in a ring of stainless steel or brass..

Oxygen Inlet Connection: Interface between cylinder valve and conserver that allows oxygen to flow into the regulator..

WARNING! Use only a manufacturer-specified seal washer.. Other seal washers may not be oxygen compatible and may cause an oxygen leak, creating an increased fire risk..

10

Image 11
Contents OM-900M Table of Contents Care and Maintenance Hygienic Preparation Intervals Figure a General InformationOperating the Unit Important Safety Rules & PrecautionsPlease remember that this is for your own safety Important Safety Rules & PRECAUTIONSACCESSORIES/REPAIRS User Qualification IntroductionPurpose Function Evolution TM Motion Conserver Description of Parts & ControlsDescription of Parts & Controls NON-PORTABLE USE Assembly and USEFigure E1 Installing Alkaline BatteriesInstalling the System Monitoring Battery Energy LevelFig.. G Proper Positioning Nasal Cannula Operating InstructionsOperating Instructions Oxygen Cylinder Duration SettingIntervals Care and MaintenanceHygienic Preparation Patient Change ProcedureDisposal Functional Check Checking for LeaksSpare Parts PRODUCTS, Spare PARTS, AccessoriesAccessories Standard ProductProblem Probable Cause Solution Troubleshooting GuideTo a user’s activity.. As a result, the device may interpret Specifications Classifications and SpecificationsSpecifications Classifications and Specifications Separation Distances Symbol Meaning Type Plate Symbols KEYLimited Warranty Important Information to Record Page EC REP