ASSEMBLY AND USE
OPERATING INSTRUCTIONS:
STEP 1: Make sure the EVOLUTIONTM MOTION conserver is set to the “PULSE” position.. Battery saving technology has eliminated the need for an “OFF” position. The unit will go to “Sleep Mode” after three minutes of
STEP 2: To reduce the risk of rapid oxygen recompression and fire, OPEN THE CYLINDER VALVE SLOWLY and completely so that the pressure gauge moves slowly as it indicates the cylinder pressure..
STEP 3: Listen for leaks.. If a leak is present, close the cylinder valve, check the seal washer, and reinstall.. If the leak persists, DO NOT USE THE EQUIPMENT.. Contact your supplier for repair..
STEP 4: To select the Rest and Active settings on the EVOLUTIONTM MOTION
the Active Selector Button on the Selector Knob until the appropriate LED display light is illuminated..
*NOTE: To check your current settings, press and release the appropriate Rest or Active Selector Button
*NOTE: The Active setting can only be set equal to or greater than that of the Rest setting.
Because the EVOLUTIONTM MOTION stores the settings from the last time it was used, you only have to complete this step before your first use or after you have made a change to your normal settings..
STEP 5: Place the nasal cannula into position with the prongs in the nostrils and begin breathing.. [Fig.. G] The EVOLUTIONTM MOTION conserver will now start to deliver oxygen.. The amount of oxygen delivered
is determined by the setting.. Adequate oxygen delivery will be achieved because of the precise time in the breathing cycle that the pulse of oxygen is delivered..
Fig.. G |
|
Proper Positioning of |
|
Nasal Cannula | 14 |
|