Evolution Technologies OM-900M manual Troubleshooting Guide, Problem Probable Cause Solution

Page 22

TROUBLESHOOTING GUIDE

PROBLEM

PROBABLE CAUSE

SOLUTION

 

Software needs to be reset

Open the battery door and remove the

 

 

batteries.. Wait 10 seconds.. Replace the

 

 

batteries and close the battery door..

 

Dead batteries.

Replace or recharge batteries.

Unit does not

Batteries installed

Make sure battery polarity is correct.

pulse.

incorrectly (reversed).

 

 

 

 

 

Dirty battery holder

Remove the batteries. Use rubbing alcohol

 

contacts.

and a cotton swab to clean contacts.

 

Cylinder valve is closed.

Turn cylinder valve clockwise to open..

 

Cylinder is empty.

Check the oxygen gauge. Replace the

 

 

cylinder, if empty.

 

 

 

 

Oxygen cannula is blocked

Remove kinks. Clean or replace, if

 

or kinked.

necessary.

 

 

 

 

Non-alkaline batteries are

Make sure the batteries inside the unit are

Short battery

used.

alkaline.

life.

Batteries are faulty.

Replace batteries. (See “Installing Alkaline

 

 

Batteries” section on page 12.)

Non-functioning units are subject to warranty provisions and the manufacturer repair/return policy.. If necessary, call your Home Care Provider..

*NOTE: Do not attempt to open the electronic compartment of the unit. If the case is opened or tampered with, the warranty is void.

21

Image 22
Contents OM-900M Table of Contents Care and Maintenance Hygienic Preparation Intervals General Information Figure aImportant Safety Rules & Precautions Operating the UnitImportant Safety Rules & PRECAUTIONS Please remember that this is for your own safetyACCESSORIES/REPAIRS Purpose IntroductionUser Qualification Function Description of Parts & Controls Evolution TM Motion ConserverDescription of Parts & Controls Assembly and USE NON-PORTABLE USEInstalling Alkaline Batteries Figure E1Monitoring Battery Energy Level Installing the SystemOperating Instructions Fig.. G Proper Positioning Nasal CannulaOperating Instructions Setting Oxygen Cylinder DurationHygienic Preparation Care and MaintenanceIntervals Procedure Patient ChangeFunctional Check Checking for Leaks DisposalStandard Product PRODUCTS, Spare PARTS, AccessoriesAccessories Spare PartsTroubleshooting Guide Problem Probable Cause SolutionTo a user’s activity.. As a result, the device may interpret Classifications and Specifications SpecificationsSpecifications Classifications and Specifications Separation Distances Symbols KEY Symbol Meaning Type PlateLimited Warranty Important Information to Record Page EC REP