Evolution Technologies OM-900M manual Procedure, Patient Change

Page 19

CARE AND MAINTENANCE

PROCEDURE

Carry out hygienic preparation of the unit and accessories as described on page 16.. We recommend using a 0..5% TERRALIN solution for wipe disinfection.. Follow the instructions enclosed with the disinfectant..

WARNING!

• Take special care that no liquids enter the unit, as this may cause damage..

• You should under no circumstances use a cleaning agent.. Cleaning agents containing alcohol or grease pose a fire risk in combination with compressed oxygen..

• Pay special attention to the oxygen inlet and outlet to make sure they remain free of dust, etc.. If the oxygen inlet connection becomes contaminated with dirt, oil, or grease, DO NOT USE OR ATTEMPT TO CLEAN. Contact your supplier for service or repair..

CAUTION!

 

• The carrying bag must never be washed in a washing machine, spin-dried or dried in a laundry

 

drier. .

 

* NOTE:

suitable gloves (e..g.. household or disposable gloves) for disinfection

You are advised to use

work. .

carrying bag, be careful not to scrub the plastic window and do not

When cleaning your

 

roughen the seams.. Repeat cleaning, if necessary.. Hang the bag in a well-ventilated area and

 

allow to air dry.. Do not hang in direct sunlight, as this may cause its external fabric to fade..

PATIENT CHANGE

Carry out a wipe disinfection on the unit’s surfaces before you hand the unit over to a new patient..

18

Image 19
Contents OM-900M Table of Contents Care and Maintenance Hygienic Preparation Intervals Figure a General InformationOperating the Unit Important Safety Rules & PrecautionsPlease remember that this is for your own safety Important Safety Rules & PRECAUTIONSACCESSORIES/REPAIRS Purpose IntroductionUser Qualification Function Evolution TM Motion Conserver Description of Parts & ControlsDescription of Parts & Controls NON-PORTABLE USE Assembly and USEFigure E1 Installing Alkaline BatteriesInstalling the System Monitoring Battery Energy LevelFig.. G Proper Positioning Nasal Cannula Operating InstructionsOperating Instructions Oxygen Cylinder Duration SettingHygienic Preparation Care and MaintenanceIntervals Patient Change ProcedureDisposal Functional Check Checking for LeaksSpare Parts PRODUCTS, Spare PARTS, AccessoriesAccessories Standard ProductProblem Probable Cause Solution Troubleshooting GuideTo a user’s activity.. As a result, the device may interpret Specifications Classifications and SpecificationsSpecifications Classifications and Specifications Separation Distances Symbol Meaning Type Plate Symbols KEYLimited Warranty Important Information to Record Page EC REP