Omron Healthcare HBF-510 instruction manual Error Indicators, Error Cause Correction Display

Page 29

ERROR INDICATORS

ERROR

CAUSE

CORRECTION

DISPLAY

 

 

 

 

 

 

 

Make sure you are barefoot.

 

 

Make sure your palms or the soles of your feet

 

Your palms or feet are not

are clean and dry.

 

Make sure your palms or feet are correctly

 

positioned over the electrodes

 

positioned.

 

correctly.

 

Refer to CORRECT POSTURE FOR

 

 

 

 

MEASUREMENT on page 22 and How to

 

 

Hold the Grip Electrodes on page 23.

 

 

 

 

The measurement position was not

Remain still and do not move until the

 

measurement is complete.

 

stable.

 

Refer to CORRECT POSTURE FOR

 

Your palms or feet were not placed

 

MEASUREMENT on page 22 and How to

 

correctly.

Hold the Grip Electrodes on page 23.

 

 

 

 

 

 

 

Slightly moisten your palms or the soles of

 

Your palms or feet are too dry.

your feet with a damp towel and retake the

 

 

measurement.

 

 

 

 

 

• Check the settings entered for your personal

 

 

data. Change the settings if needed. Retake

 

 

the measurement.

 

The Body Composition values are

• The main unit cannot measure body

 

outside the measurement range.

composition outside its measurement range,

 

 

even if the age, gender and height settings are

 

 

correct. Refer to Display and Set Ranges on

 

 

page 34.

 

 

 

 

 

Remove the batteries. Wait one minute.

 

 

Reinstall the batteries. Refer to BATTERY

 

The monitor is not set up correctly.

INSTALLATION on page 16.

 

 

Press the Power Switch and turn the monitor

 

 

on. Retake the measurement.

 

 

 

 

You stepped onto the measurement

Wait until 0.0 lb is displayed before stepping

 

platform before 0.0 lb was

onto the measurement platform.

 

displayed.

 

 

 

 

 

 

The monitor was moved before

Do not move the monitor until 0.0 lb is

 

0.0 lb was displayed.

displayed.

 

 

 

 

Movement during measurement.

Do not move until measurement is complete.

 

 

 

 

Monitor detected weight over

This monitor can only be used by people

 

330.0 lb.

weighing less than 330.0 lb.

 

 

 

29

Image 29
Contents Full Body Sensor Body Composition Monitor Scale Table of Contents Date Purchased Serial Number IntroductionOperating the Device Safety InformationCare and Maintenance Principles of Full Body Sensing Calculations Information on Body CompositionBody FAT Body Mass IndexNormal + High ++ Very High Visceral FATSkeletal Muscle 170 160 150 LONG-TERM Planning for Successful Weight LossAvoid Taking Measurements Under the Following Conditions Recommended Measurement TimesResults may differ from actual body fat percentage Display Unit Know Your UnitKnow Your Unit Power Switch Using the Display UnitLocate the Battery Cover on the back Battery InstallationInstall 4 AA size batteries so the + Replace the battery coverBattery Replacement Lift the Display Unit out of the Display Unit Holder Setting Personal DataPress the Power Switch to turn the monitor on Enter your Personal Data Select a Personal Profile Number or the Guest modeChanging Personal Data Setting the GenderSelect the Personal Profile Number you want to change Deleting Personal DataRefer to on pages 19 and 20 to enter your personal Data Select the Personal Profile Number you want to deleteCorrect Posture for Measurement HOW to Take a MeasurementHow to Hold the Grip Electrodes HOW to Measure Body CompositionPress the Guest button while holding the Display Unit Start the measurement Press the SET/MODE button to check the measurement results Understanding Your Measurement Results Check the measurement result HOW to Measure Weight onlyStep onto the Measurement Platform Error Cause Correction Display Error IndicatorsProblem Cause and Solution Troubleshooting TipsCare and Maintenance FCC Statement Limited Warranty WarrantyBMI Specifications