EXCELL PRECISION CO., LTD.
F. How to set the speed of sampling and zero range
The speed of sampling:
The speed setting is from 01~15. The smaller number you set, the slower speed for sampling is. The slower speed you set, the scale will show more accurate. If faster speed you set, the scale will sacrifice the accuracy.
Factory setting for sampling is ⇒ 01
Zero display range:
Zero display range: 0 ⇒ Normal standard status
1 ⇒ The 1st division hidden, not appeared
Factory setting for zero range ⇒ 1
(1) Press
→0←
key followed by 8 key.
XX
Please use 6 or 4 key to input “0” or “1” .
(0)(1)
Please use 8 or 2 key to choose the number 01~15 of speed you need.
(#) ($)
(2)Press
key, the scale will return to the counting mode.
→0← 3
.
Stable 3
27