When the control solution test result is within the control range, you are ready to test your blood. Your con- trol solution test will be stored in the meter memory. Record control solution tests in your logbook by placing a C next to the result to indicate a control solution test.
| NOTES |
TEST |
|
DATE |
|
TEST |
|
RESULT |
|
TIME | NOTES |
INSULIN/ |
|
MEDICATION |
|
TEST |
|
DATE |
|
TEST |
|
RESULT |
|
TIME |
|
INSULIN/ |
|
MEDICATION |
|
Control Solution Range
If test results fall outside the expected range, repeat the test. Results that fall outside the expected range may indicate the system is not working properly. Some possible causes are:
■Procedural error
■Expired or contaminated control solution
■Test strip code and meter code do not match
■Dirty meter or test strip holder
■Test strip deterioration
■Meter malfunction
■Control solution that is outside of the temperature range
If you are unable to resolve the problem, call LifeScan Customer Services, 1 800
CHECKING