Lifescan blood glucose monitoring system manual Blood Application

Page 76

BLOOD APPLICATION

How you can be sure that you have applied the right amount of blood:

REFERENCE

Front Back

Pink

Test

Square

White Confirmation

Pad Dot

Enough Blood

Front Back

Not Enough

Blood

Back

Too Much Blood*

Front

QUICK

Pink test square is where you apply a drop of blood.

White pad absorbs the excess blood that may extend beyond the pink test square. Do not apply blood directly onto the white pad.

Confirmation dot on back

of test strip is where you check to confirm proper blood application.

Apply a drop of blood to the pink test square on the front of the test strip. The confirmation dot on the back should be completely blue for an accurate test.

If any white patches or streaks appear on the blue confirmation dot, you have not applied enough blood for an accurate test. You must start the test over with a new test strip to avoid false low results.

If there is no white visible on the white pad, you have applied too much blood. You must start the test over with a new test strip to avoid false high results.

*Be careful not to apply too much blood, especially if you tend to bleed easily or are taking blood-thinning medications.

68

Image 76
Contents Owner’s Booklet Page Important phone numbers Meter Serial Number Page TA B L E O F C O N T E N T S Checking the SystemUnderstanding Your Results Solving Problems Page A R N I N G T H E S Y S T E M Quick Reference Surestep Blood Glucose Meter Holds two AAA alkaline batteriesSurestep Test Strips Black Tip Confirmation DotImportant Test Strip Information Display Symbols Example Test Result HI mg/dL Clean Meter and Test Strip Holder Replace the Two AAA Alkaline BatteriesD I N G T H E M E T E R Three Easy Steps to CodingCheck the Code Numbers Code the MeterSurestep Glucose Control Solution E C K I N G T H E S Y S T E MImportant Control Solution Test Information Before Testing, Check the Code Apply a Small Drop of Control SolutionCheck Confirmation Dot Insert the Test Strip and Read ResultSurestep Control Solution Range Getting a Drop of Blood S T I N G Y O U R B L O O DInsert a Lancet Penlet II Automatic Blood Sampler Testing Cock the Penlet II Sampler Get a Drop of Blood Testing Remove the Lancet Step-by-Step Test Procedure Before Testing On the display Apply Blood to the Test StripTesting Too Much Blood Check the Confirmation DotNot Enough Blood Example Kahn, R., and Weir, G. Joslin’s Diabetes Mellitus, 13th Optional Color Chart Color Chart is not a replacement for a Surestep Meter testTesting D E R S TA N D I N G Y O U R R E S U LT S OverviewRecent Diabetes Findings What This Means for YouFrequency of Testing Comparing Meter Laboratory ResultsExpected Test Results Target Range Glucose Ranges For People Without Your Target Testing Tips to Ensure Accuracy After TestingHealth Conditions Causing Out-of-Range Results Results Recommendations for Better Diabetes Control ReferencesUsing the Meter for the First Time Recalling Results from Meter MemoryRecalling All Results in Meter Memory Recalling Your Last Result R I N G F O R Y O U R M E T E R Cleaning the MeterGeneral Care Clean the Test Strip Holder Remove the Test Strip HolderClean the Lens Area and Contact Points Replace the Test Strip Holder Changing the Batteries Remove the Battery Door Remove the Old Batteries Close the Battery DoorInsert New Batteries A N G I N G M E T E R S E T T I N G S Set-Up Mode Entering Set-Up Mode Display of Last ResultUnit of Measurement Beep Signals Settings LV I N G P R O B L E M S Helpful MessagesFacing up Error Messages Error Possible Cause What to doConfirmation dot versus Error Possible Cause What to do Error Possible Cause What to do Error Possible Cause What to do Performance Characteristics E C I F I C AT I O N SDay Money-Back Guarantee Three-Year WarrantyWarranty Coding the Meter Blood Application A. English 1 800 Español 1 800
Related manuals
Manual 103 pages 12.78 Kb Manual 85 pages 3.35 Kb