PROGRAM INSTRUCTIONS (cont)
7.Depress MODE (S1) switch once.
The text �� followed by the current value of Hysteresis Low will be displayed. Repeat Step 2 to set Alarm Hysteresis Low to the desired value. The maximum value for HL is 99 counts. Proceed to Step 8 or depress S4 to exit.
8.Depress MODE (S1) switch once.
The text ��� followed by the current CAL number will be displayed. Use S2 & S3 to set the Cal Number. Remember the Cal Value should be between 2000 & 19000 (decimal not used). Proceed to Step 9 or depress S4 to exit.
9. Depress MODE (S1) switch once.
The text ���� followed by the current decimal position will be shown flashing. Use S2 to set the decimal point to the desired location.
10.Depress MODE (S1) switch once.
The instrument will exit the Mode Selection Sequence and enter the normal operating mode (i.e. monitoring the input signal) This is the end of the PROGRAM INSTRUCTIONS.
Page 12