Sierra 951 Alarm Set Point High 2 SPH2, Alarm Set Point Low 2 SPL2, Alarm Hysteresis High HH

Page 9

FUNCTION DEFINITIONS (cont)

Alarm Set Point High 2 (SPH2)

SPH2 is independent and functions identically to SPH1. SPH2 has its own open collector outputs and front panel LED (H2).

Alarm Set Point Low 2 (SPL2)

SPL2 is independent and functions identically to SPL1. SPL2 has its own open collector outputs and front panel LED (L2).

Alarm Hysteresis High (HH)

HHis the hysteresis value for the alarms SPH1 and SPH2. The hysteresis value determines the number of counts the displayed reading must fall below SPH1 and SPH2 values before deactivating their respective open collector outputs and LEDs. The maximum value of hysteresis is 99 counts while the minimum is 00 counts.

Alarm Hysteresis Low (HL)

HL is the hysteresis value for the alarms SPL1 and SPL2. The hysteresis value determines the number of counts the displayed reading must exceed the SPL1 and SPL2 values before deactivating their respective open collector outputs and LEDs. The maximum value of hysteresis is 99 counts while the minimum is 00 counts.

FS Cal (CAL)

This input allows the user to calibrate the full scale readout to any desired engineering units. This input must be entered prior to performing the calibration. The FS Cal number may be any number from 1 to 99,999. Zero is an invalid FS Cal number. For best resolution and stability, choose a CAL number between 2000 and 19000 counts.

Example: If the Flowmeter is to be calibrated at a full scale flowrate of

500 sccm, the FS Cal value could be 500.00, 500.0 or 500. If the FS Cal value of 500.00 is used, the unit will automatically count by 3 when full scale calibration is completed since the analog-to-digital converter resolution is only 19,999 counts. Choosing a FS Cal value of 500.0 or 500 assures the best resolution and stability in this particular case.

Page 8

Image 9
Contents Sierra Model Digital Flo-Box Description SpecificationsInstallation and Wiring InstallationWiring Annunciator leds 15 vdc Switch Definitions Switch and Function DefinitionsMenu Switch S1 Increment Switch S2Step and Min-Max Switch S3 Zero and Exit Switch S4Cal Switch S5 Alarm Set Point Low 1 SPL1 Alarm Set Point High 1 SPH1Function Definitions Flow Set Point StPtAlarm Set Point Low 2 SPL2 Alarm Set Point High 2 SPH2Alarm Hysteresis High HH Alarm Hysteresis Low HLDecimal Point dP Operating Instructions TURN-ON InstructionsS1 S2 S3 S4 S5 Page Page Calibration Instructions Page Special Order Only RS232 Serial CommunicationsSerial Lockout FS CAL Model SR5 Program Menu Diagram