CALIBRATION INSTRUCTIONS (cont)

6.Depress and hold CAL (S5) for approximately 3 seconds. The FS CAL LED should illuminate and the readout should display the FS Cal value within +/- 1 digit.

7.The CALIBRATION sequence is now complete.

NOTE: Calibration information (zero and scale factor values) are stored in non-volatile memory and are updated every time calibration step 6 above is performed. ZEROing the display changes and stores a new zero value but does not affect the scale factor value.

Page 14

Image 15
Contents Sierra Model Digital Flo-Box Description SpecificationsInstallation and Wiring InstallationWiring Annunciator leds 15 vdc Increment Switch S2 Switch and Function DefinitionsSwitch Definitions Menu Switch S1Step and Min-Max Switch S3 Zero and Exit Switch S4Cal Switch S5 Flow Set Point StPt Alarm Set Point High 1 SPH1Alarm Set Point Low 1 SPL1 Function DefinitionsAlarm Hysteresis Low HL Alarm Set Point High 2 SPH2Alarm Set Point Low 2 SPL2 Alarm Hysteresis High HHDecimal Point dP Operating Instructions TURN-ON InstructionsS1 S2 S3 S4 S5 Page Page Calibration Instructions Page Special Order Only RS232 Serial CommunicationsSerial Lockout FS CAL Model SR5 Program Menu Diagram