MAINTENANCE SCHEDULE

Info: Only qualified persons should service and repair your Lumex patient lift. Regular maintenance of your patient lift is necessary to ensure continuing proper and safe opera- tion. Read and observe the following recommended maintenance schedule:

Item

Inspect

 

Inspect/adjust

 

initially

 

 

 

 

 

daily

 

monthly

every 6

 

 

 

 

 

 

 

 

months

 

 

 

 

 

 

Base and Casters

 

 

 

 

 

 

 

 

 

 

 

Ensure that base is level (all four wheels touch the floor)

X

 

 

X

X

Ensure that caster mounting bolts and axle bolts are tight

X

 

 

X

 

Ensure that casters swivel and roll freely

X

 

 

X

 

Apply a light grease to caster ball bearings

 

 

 

 

X

Base Adjustments

 

 

 

 

 

 

 

 

 

 

 

Ensure that base opens and closes smoothly

X

 

 

X

X

Ensure that base locks securely in three positions

X

 

 

X

 

Ensure that locking knob is in place and operating properly

X

 

 

X

 

Inspect all base hardware for wear

 

 

 

X

 

Mast

 

 

 

 

 

 

 

 

 

 

 

Ensure that mast is fully engaged and locked into mast sleeve

X

 

 

X

 

Inspect mast for bends or deflections

 

 

 

X

 

Inspect all mast hardware for wear

 

 

 

X

X

Boom

 

 

 

 

 

 

 

 

 

 

 

Inspect boom for bends or deflections

X

 

 

X

 

Ensure that boom is centered between the base legs

X

 

 

X

X

Inspect all boom hardware for wear

 

 

 

X

 

Spreader Bar

 

 

 

 

 

 

 

 

 

 

 

Inspect hanger post for wear or damage

 

 

 

X

 

Inspect sling hook for wear or deflections

 

 

 

X

X

Inspect all spreader bar hardware for wear

 

 

 

X

X

Slings

 

 

 

 

 

 

 

 

 

 

 

Inspect slings and straps for wear or damage

 

X

 

 

 

Inspect metal/plastic parts for wear or damage

 

X

 

 

 

Lubrication with light oil

 

 

 

 

 

 

 

 

 

 

 

Lubricate all pivot points

 

 

 

 

X

Lubricate top and bottom of actuator mounting

 

 

 

 

X

Cleaning

 

 

 

 

 

 

 

 

 

 

 

Clean with mild non-abrasive detergent

 

 

 

 

X

Pump

 

 

 

 

 

Ensure that pump mounting hardware is tight

X

 

 

X

 

Inspect pump mounting hardware for wear

 

 

 

X

 

1031FP-INS-LAB-RevA11

18

Image 18
Contents LF1031FP Hydraulic Patient Lift With Foot Pedal Contents Introduction Safety Precautions 1031FP-INS-LAB-RevA11 Features Unpacking Handling ProceduresShipping and Transportation Instructions InspectionAssembly Instructions ComponentsAssembly Install push handles Install spreader bar Lifting the patient from a bed Operating InstructionsLifting the Patient Lowering the patient onto a bedLifting the patient from the floor Lowering the patient to the floorRemoving the sling Care and Maintenance Before Each USELubrication Cleaning Checking for Damage and WearCare of Slings Installation of 1010S005 Hydraulic Pump Removal of Existing PumpInstallation of New Pump Maintenance Schedule Troubleshooting Symptom Possible Cause ActionSpecifications Limited Warranty GF Health Products, Inc. Graham-Field Index