MAINTENANCE SCHEDULE
Info: Only qualified persons should service and repair your Lumex patient lift. Regular maintenance of your patient lift is necessary to ensure continuing proper and safe opera- tion. Read and observe the following recommended maintenance schedule:
Item | Inspect |
| Inspect/adjust | ||
| initially |
|
|
|
|
| daily |
| monthly | every 6 | |
|
|
| |||
|
|
|
|
| months |
|
|
|
|
|
|
Base and Casters |
|
|
|
|
|
|
|
|
|
|
|
Ensure that base is level (all four wheels touch the floor) | X |
|
| X | X |
Ensure that caster mounting bolts and axle bolts are tight | X |
|
| X |
|
Ensure that casters swivel and roll freely | X |
|
| X |
|
Apply a light grease to caster ball bearings |
|
|
|
| X |
Base Adjustments |
|
|
|
|
|
|
|
|
|
|
|
Ensure that base opens and closes smoothly | X |
|
| X | X |
Ensure that base locks securely in three positions | X |
|
| X |
|
Ensure that locking knob is in place and operating properly | X |
|
| X |
|
Inspect all base hardware for wear |
|
|
| X |
|
Mast |
|
|
|
|
|
|
|
|
|
|
|
Ensure that mast is fully engaged and locked into mast sleeve | X |
|
| X |
|
Inspect mast for bends or deflections |
|
|
| X |
|
Inspect all mast hardware for wear |
|
|
| X | X |
Boom |
|
|
|
|
|
|
|
|
|
|
|
Inspect boom for bends or deflections | X |
|
| X |
|
Ensure that boom is centered between the base legs | X |
|
| X | X |
Inspect all boom hardware for wear |
|
|
| X |
|
Spreader Bar |
|
|
|
|
|
|
|
|
|
|
|
Inspect hanger post for wear or damage |
|
|
| X |
|
Inspect sling hook for wear or deflections |
|
|
| X | X |
Inspect all spreader bar hardware for wear |
|
|
| X | X |
Slings |
|
|
|
|
|
|
|
|
|
|
|
Inspect slings and straps for wear or damage |
| X |
|
|
|
Inspect metal/plastic parts for wear or damage |
| X |
|
|
|
Lubrication with light oil |
|
|
|
|
|
|
|
|
|
|
|
Lubricate all pivot points |
|
|
|
| X |
Lubricate top and bottom of actuator mounting |
|
|
|
| X |
Cleaning |
|
|
|
|
|
|
|
|
|
|
|
Clean with mild |
|
|
|
| X |
Pump |
|
|
|
|
|
Ensure that pump mounting hardware is tight | X |
|
| X |
|
Inspect pump mounting hardware for wear |
|
|
| X |
|