Graham Field LF1031FP user manual Troubleshooting, Symptom Possible Cause Action

Page 19

8TROUBLESHOOTING

WARNING: If you experience a problem with your lift and are unable to service it your- self, contact your Lumex dealer.

SYMPTOM

POSSIBLE CAUSE

ACTION

Patient lift does not go up

Air in the cylinder

Bleed the cylinder by turning the release

or down

 

knob and push down on the mast. Pump

 

 

up the lift again and turn the release knob

 

 

again.

 

Loss of hydraulic fluid (leaking

Replace any leaking cylinder

 

cylinder)

 

Patient lift does not go

Air in the cylinder

See above Directions

down

 

 

Info: If the hydraulic cylinder begins leaking fluid, do not operate. Contact your Lumex dealer for a new cylinder.

1031FP-INS-LAB-RevA11

19

Image 19
Contents LF1031FP Hydraulic Patient Lift With Foot Pedal Contents Introduction Safety Precautions 1031FP-INS-LAB-RevA11 Features Inspection Handling ProceduresShipping and Transportation Instructions UnpackingComponents Assembly InstructionsAssembly Install push handles Install spreader bar Lowering the patient onto a bed Operating InstructionsLifting the Patient Lifting the patient from a bedLowering the patient to the floor Lifting the patient from the floorRemoving the sling Before Each USE Care and MaintenanceLubrication Checking for Damage and Wear CleaningCare of Slings Removal of Existing Pump Installation of 1010S005 Hydraulic PumpInstallation of New Pump Maintenance Schedule Symptom Possible Cause Action TroubleshootingSpecifications Limited Warranty GF Health Products, Inc. Graham-Field Index