2.4 Signal connection

a) Connection of video signals

The connected equipment must comply to all relevant safety demands.

Important: The use of low-quality video cables can distort the video signal and influence diagnosis.

1.Check the impedance of the imaging board that produces the video signals you want to connect. It must be 75 Ohm (just like the input of the display). If not, the quality of the images on the display will be inferior.

2.Connect the video output of the computer’s imaging board to the video and sync inputs on the display’s rear panel. Use a proper video cable. The video cable is not supplied with the display, unless the display comes as part of a complete BARCO MeDis system, that also contains an imaging board.

The inputs accept the following signals:

Video with separate horizontal and vertical sync.

-The video cable has 3 wires.

-Connect the video signal to the connector Video {7}.

-Connect the horizontal sync signal to the connector HS/CS {8}.

-Connect the vertical sync signal to the connector VS {9}.

Video with external composite sync.

-The video cable has 2 wires.

-Connect the video signal to the connector Video {7}.

-Connect the composite sync signal to the connector HS/CS {8}.

Video with internal composite sync (sync on video).

-The video cable has 1 wire.

-Connect the video (with sync) signal to the connector Video {7}. Notes:

The video inputs cannot be connected in loop-through (daisy- chain).

The required video amplitude: 700 mV ± 3 dB.

The required sync. amplitude: 500 mV.

Installation13

Image 12
Contents MGD 221 Mkii Medical Greyscale DisplayTrademark information MGD 221 Users Guide Safety Instructions MGD 221 Users Guide Contents MGD 221 Users Guide Introduction OverviewPower saving system Versions and optionsAbout the manuals Installation PrecautionsPackage contents Controls and connectors Signal connection Connection of video signalsConnection of data signals Connection of optical sensorConnection of power Positioning the displayImportant considerations Switching on / off Power on/offAutomatic power saving system Dpms Operation User ControlsContrast and brightness control To activate the OSDTo put contrast in calibrated position To control contrast manually To put brightness in calibrated positionTo control brightness manually Viewing Status information To view the status informationExplanation of the Status information menu items Other adjustments Activity timeoutAdvanced Settings and Adjustments For Service Staff only SettingsTo change the settings Explanation of the Settings menu items Saving the settingsAdjustments How to make the adjustmentsDescription of the adjustments Geometry RotationAdvanced Interlacing Pre correction Interlacing Post correctionSH Compensation optional SH Compensation SH Compensation ON/OFFSH Compensation SH Position Advanced Advanced To adjust the ALC system, proceed as follows Focus Static focusFocus zone ALC optionalFactory reset How to save changes Proceed as followsLuminance uniformity What happens upon saving?Maintenance Picture tubeCabinet Troubleshooting There appears no image, the green LED is onThere appears no image, the green LED is blinking Appendix a Background Information Luminance Uniformity CorrectionPage Appendix B Technical Specifications AR panel 30%