Barco MGD 221 MKII user manual Viewing Status information, To view the status information

Page 19

4Turn the control knob until you have reached the desired brightness.

5Press the control knob shortly or wait a few seconds to return to the brightness control menu.

6In the brightness control menu, select Exit and press the control knob shortly to return to the main menu. Alternatively, you can just wait a few seconds, and the OSD returns to the previous menu automatically.

3.3Viewing Status information

You can view the status and actual settings of the display. You cannot change them in the Status Menu. Changing the settings is reserved for trained service staff only.

To view the status information

Turn the control knob to select Status information in the OSD main menu.

The Status information menu appears. The only control in the menu is Exit.

STATUS INFORMATION

Display Name:

MGD 221MKII

Serial Number:

5114147

Mode Name:

New mode

Mode H-frequency: 48.32 kHz

Mode V-frequency:

59.94 Hz

Software Version:

V 1.01

Run time:

50 hours

White level:

215 cd/m2

Activity timeout

30s

SH compensation

OFF

Actual

H-frequency:72.02 kHz

Actual

V-frequency: 75.00 Hz

- Exit

Status information menu

Explanation of the Status information menu items

Display Name:

This is the display type.

Serial Number:

This is the display's serial number.

Mode Name:

The currently selected scanning mode.

Mode H-frequency, Mode V-frequency:

These are the horizontal and vertical sync frequencies of the currently selected scanning mode. These are the figures stored in the memory.

Operation20

Image 19
Contents Medical Greyscale Display MGD 221 MkiiTrademark information MGD 221 Users Guide Safety Instructions MGD 221 Users Guide Contents MGD 221 Users Guide Overview IntroductionVersions and options Power saving systemAbout the manuals Precautions InstallationPackage contents Controls and connectors Connection of video signals Signal connectionConnection of optical sensor Connection of data signalsPositioning the display Connection of powerImportant considerations Operation User Controls Switching on / offPower on/off Automatic power saving system DpmsTo activate the OSD Contrast and brightness controlTo put contrast in calibrated position To put brightness in calibrated position To control contrast manuallyTo control brightness manually To view the status information Viewing Status informationExplanation of the Status information menu items Activity timeout Other adjustmentsSettings Advanced Settings and Adjustments For Service Staff onlyTo change the settings Saving the settings Explanation of the Settings menu itemsGeometry Rotation AdjustmentsHow to make the adjustments Description of the adjustmentsAdvanced SH Compensation SH Compensation ON/OFF Interlacing Pre correctionInterlacing Post correction SH Compensation optionalSH Compensation SH Position Advanced Advanced ALC optional To adjust the ALC system, proceed as followsFocus Static focus Focus zoneWhat happens upon saving? Factory resetHow to save changes Proceed as follows Luminance uniformityPicture tube MaintenanceCabinet There appears no image, the green LED is on TroubleshootingThere appears no image, the green LED is blinking Luminance Uniformity Correction Appendix a Background InformationPage AR panel 30% Appendix B Technical Specifications