Fluke 41B, 39 user manual Screen Abbreviations, Screen Usage Name and Description

Page 32

39/41B Users Manual

 

Table 3. Screen Abbreviations (cont)

 

 

 

Screen Usage

 

Name and Description

 

 

 

DPF

 

Displacement Power Factor cosφ fundamental

 

 

The ratio of the active power (W) to the apparent power

 

 

(VA rms) at the fundamental frequency. Equivalent to

 

 

cosφ at that frequency. Traditional power factor for linear

 

 

loads.

HZ

 

Waveform Screens: Fundamental frequency in hertz

 

 

 

Harmonics Screens: Frequency of selected harmonic in

 

 

hertz

KF

 

K-Factor

 

 

 

A load current waveform rating calculation that rates the

 

 

waveform’s capability to produce harmonic-related heat

 

 

loss in transformers and other magnetic components.

 

 

Used to select transformer K-ratings that match load

 

 

current K-factor measurements.

PF

 

Power Factor

 

 

 

Ratio of active power to apparent power (including all

 

 

harmonics).True power factor for all loads, linear and

 

 

non-linear.

 

 

 

%THD-FTotal Harmonic Distortion (as % of Fundamental)

Defines amount of harmonic distortion as a percentage of the waveform at the fundamental frequency.

%THD - F = rms of harmonics (less fundamental) rms of fundamental

%THD-RTotal Harmonic Distortion (as % of rms total)

Defines amount of harmonic distortion as a percentage of the rms value of waveforms at all frequencies (fundamental and harmonics).

%THD - R = rms of harmonics (less fundamental) total rms

22

Image 32
Contents Fluke 39/41B Limited Warranty & Limitation of Liability Service Centers Page Table of Contents Fluke 39/41B List of Tables List of Figures Using!Your Tester Safely AC Current Probe Safety Guidelines About This Manual Uses for the TesterUnpacking Unpacking Additional Items Model 41B Terminology Used in This Manual Memory Retention Battery ConsiderationsAutomatic Power Down Gettingto Acquainted with Your Tester Getting Acquainted Using the Voltage Test Leads Input ConnectionsUsing the Current Probe Measurement Connections 1PThree-Phase Power Measurements Three-Phase Measurement Connections For Balanced CircuitsThree-Phase Measurement, Balanced or Unbalanced Wtotal = W1 + W2Wtotal = W1 + W2 + W3 Keypad KeypadNumber Key Descriptions Symbol Name & Description KeypadNumber Symbol Name & Description Key Descriptions Number Name & Description Key Descriptions Number Symbol Name & Description Print Model 41BDisplay Screen Display ScreenStatus Line Screen Abbreviations Screen Usage Name and Description DP ase Angle degrees. ±Screen Abbreviations Screen Usage Name and DescriptionVolts rms includes dc component Special Messages Waveform Screen Mode Functions and Screen ModesWaveform Screens Functions and Screen Modes Harmonics Screens Harmonics Overload Screen Harmonics Zoom Mode Harmonics Zoom ScreenText Screen Mode Text Screens VA Check Screen Typical VA Check Signatures Recording Measurements Record Screens Power-Up Configuration THD RMSConfiguration Screens Storing & Recalling Measurements Model 41B Memory Screen39/41B Store Using the Serial Cable Model 41B Serial Communications Model 41B Sending to a Printer or PC Model 41B Using PC Software with the Tester Model 41BPrinting Model 41B Applications Applications Plant Switch Gear Service Entrance Plant Switch GearDistribution Transformer Distribution Transformer 1PWatts Text ScreenElectronic Equipment Load Center Distribution Transformer 3P, 3C Watts Text ScreenReceptacle Load Center Neutral Current Measurement Generator Set Generator SetAdjustable Frequency Speed Motor Drive Adjustable Frequency Motor DriveUser Maintenance CleaningFuse Replacement Battery ReplacementUser Maintenance If Your Tester Does Not Work Accessories and Replacement Parts Accessories and Replacement PartsQty Replacement Parts DescriptionFluke PN Specifications MV/A Isolated InputHarmonic Level 5% Using Smooth ~20 Watts Measurements Volt-AmpsAccuracy Range/ResolutionMeasurement Function Watts AC VoltsAC Amps General Specifications Display Safety 39/41B