Fluke 39, 41B user manual Recording Measurements

Page 45

R

Recording Measurements

Recording Measurements

Press Rto begin recording readings for all measurement functions. Then press <or >to cycle through the screens for the function presently selected. You can also switch between measurement functions while in Record mode. (See Figure 16 for an overview of possible record screens.) For any function, the Tester displays screens in the following order:

NOW The present readings. This screen always appears first when you start Record.

MAX The maximum values measured since you started Record.

AVG The averages of values measured since you started

Record. (Average values stop updating after 48 hours of continuous record operation.)

MIN The minimum values measured since you started Record. WhenE<>in ecord, only the following keys are active:C

F

Press Ra second time to exit Record. This action discards all recorded values and returns to the previous measurement screen. The Tester begins storing a new set of values each time you start Record.

The Tester loses recorded values if the batteries discharge too much. If b comes on during Record, write down any values of interest. (You have ample time to do this with alkaline batteries. NiCad batteries discharge much more rapidly once bcomes on.) Changing the battery erases the Record memory.

You can access 12 different record screens: NOW, MAX, AVG, and MIN for volts, amps, and watts measurement functions. Figure 16 presents an overview of the record measurements and computations available by function.

Note

The Record mode acquires only single-phase data approximately every 500 ms (0.5s). It is not intended to capture transients, nor is it able to capture single events on the power line.

35

Image 45
Contents Fluke 39/41B Limited Warranty & Limitation of Liability Service Centers Page Table of Contents Fluke 39/41B List of Tables List of Figures Using!Your Tester Safely AC Current Probe Safety Guidelines Uses for the Tester About This ManualUnpacking Unpacking Additional Items Model 41B Terminology Used in This Manual Battery Considerations Automatic Power DownMemory Retention Gettingto Acquainted with Your Tester Getting Acquainted Input Connections Using the Voltage Test LeadsMeasurement Connections 1P Using the Current ProbeThree-Phase Measurement Connections For Balanced Circuits Three-Phase Power MeasurementsWtotal = W1 + W2 Three-Phase Measurement, Balanced or UnbalancedWtotal = W1 + W2 + W3 Keypad KeypadKeypad Number Key Descriptions Symbol Name & DescriptionNumber Symbol Name & Description Key Descriptions Number Name & Description Print Model 41B Key Descriptions Number Symbol Name & DescriptionDisplay Screen Display ScreenStatus Line DP ase Angle degrees. ± Screen Abbreviations Screen Usage Name and DescriptionScreen Usage Name and Description Screen AbbreviationsVolts rms includes dc component Special Messages Functions and Screen Modes Waveform Screen ModeWaveform Screens Functions and Screen Modes Harmonics Screens Harmonics Overload Screen Harmonics Zoom Screen Harmonics Zoom ModeText Screen Mode Text Screens VA Check Screen Typical VA Check Signatures Recording Measurements Record Screens THD RMS Power-Up ConfigurationConfiguration Screens Memory Screen Storing & Recalling Measurements Model 41B39/41B Store Using the Serial Cable Model 41B Serial Communications Model 41B Using PC Software with the Tester Model 41B Printing Model 41BSending to a Printer or PC Model 41B Applications Applications Plant Switch Gear Plant Switch Gear Service EntranceDistribution Transformer 1PWatts Text Screen Distribution TransformerDistribution Transformer 3P, 3C Watts Text Screen Electronic Equipment Load CenterReceptacle Load Center Neutral Current Measurement Generator Set Generator SetAdjustable Frequency Motor Drive Adjustable Frequency Speed Motor DriveCleaning User MaintenanceFuse Replacement Battery ReplacementUser Maintenance If Your Tester Does Not Work Accessories and Replacement Parts Accessories and Replacement PartsReplacement Parts Description Fluke PNQty MV/A Isolated Input SpecificationsWatts Measurements Volt-Amps Harmonic Level 5% Using Smooth ~20Range/Resolution Measurement FunctionAccuracy AC Volts AC AmpsWatts General Specifications Display Safety 39/41B