Leica 760498a manual Appendix, Message codes, Technical data

Page 19

Appendix

Message codes

All message codes are displayed with either or "Error". The following errors can be corrected:

 

Cause

Remedy

156

Transverse tilt

Hold the instrument

 

greater than 10°

without any transverse

 

 

tilt

160

Main tilt direction,

Measure angle up to

 

angle too high

max. ± 45°

 

(> 45°)

 

204

Calculation error

Repeat procedure

252

Temperature too

Cool down instrument

 

high

 

253

Temperature too

Warm up instrument

 

low

 

255

Receiver signal too

Use target plate

 

weak, measurement

 

 

time too long,

 

 

distance > 100 m

 

256

Received signal too

Target too reflective

 

strong

(use target plate)

257Wrong measure- Darken target (measure

ment, background

in different lighting

brightness too high

conditions)

260 Laser beam inter-

Repeat measurement

rupted

 

Error

Cause

Remedy

Error

Hardware error

Switch on/off the

 

 

device several times. If

 

 

the symbol still

 

 

appears, then your

 

 

instrument is defective.

 

 

Please call your dealer

 

 

for assistance.

 

 

 

Technical data

Distance

typically:

measurements:

Measuring accuracy

± 1.5 mm*

up to 30 m

± 1/16 ’’*

(2 σ, standard deviation)

 

Power Range

 

Technology™:

0.05 m to 100 m

Range (use target plate

 

from about 80m)

 

Smallest unit displayed

0.1 mm

Distance measurement

9

Minimum/maximum

 

measurement, Continuous

9

measurement

 

Area/volume calculation of

9

room data

 

Addition / subtraction

9

Indirect measurement

9

using Pythagoras

 

EN

Leica DISTO™ D3 760498a en

15

Appendix

Image 19
Contents Leica Distotm D3 Leica Disto D3 10 11 12 1314 15 16,17,18 Page Use of the instrument ContentsSafety Instructions Symbols usedAreas of responsibility Limits of useElectromagnetic Compatibility Hazards in useLaser classification FCC statement applic. in U.SMultifunctional endpiece Start-upLabelling Inserting/replacing batteriesDisplay KeypadMenu functions SettingsSetting the unit for distance mea Surements Setting the unit for tilt measure- mentsMeasuring with the tripod Tripod Switching on and offOperation Continuous laserSingle distance measurement Reference settingMeasuring Display / keypad illuminationArea VolumeFunctions Addition / subtractionRoom corner angle function Direct horizontal distanceStake out function Indirect measurement Functions Storage of constants/historical storage Timer self-triggeringAppendix Technical dataMessage codes Tilt measurements Measuring conditionsCare WarrantyPage Page Exit Leica Disto D3