Status Subsystem

Status Subsystem

This subsystem programs the power supply status registers. The power supply has three groups of status registers; Operation, Questionable, and Standard Event. The Standard Event group is programmed with Common commands as described in "Chapter 4 - Status Reporting". The Operation and Questionable status groups each consist of the Condition, Enable, and Event registers and the NTR and PTR filters.

STAT:PRES

This command sets all defined bits in the Status Subsystem PTR registers and clears all bits in the subsystem NTR and Enable registers. STAT:OPER:PTR is set to 1313 and STAT:QUES:PTR is set to 1555.

Command Syntax

STATus:PRESet

Parameters

(None)

Examples

STAT:PRES STATUS:PRESET

Query Syntax

(None)

Related Commands

(None)

Status Operation Registers

The bit configuration of all Status Operation registers is shown in the following table:

Bit Configuration of Operation Registers

 

Bit Position

15-12

11

10

9

8

7

6

5

4

3

2

1

0

 

 

Bit Name

NU

NU

CC

NU

CV

NU

NU

WTG

NU

NU

NU

NU

CAL

 

 

Bit Weight

 

2048

1024

512

256

128

64

32

16

8

4

2

1

 

 

CAL = Interface is computing new calibration constants; CC = The power module is in constant current mode.

 

 

 

CV = The power module is in constant voltage mode; NU = (Not used); WTG = Interface is waiting for a trigger.

 

 

 

 

 

 

 

 

 

Note

See "Chapter 4 - Status Reporting" for more explanation of these registers.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

STAT:OPER?

This query returns the value of the Operation Event register. The Event register is a read-only register which holds (latches) all events that are passed by the Operation NTR and/or PTR filter. Reading the Operation Event register clears it.

Query Syntax

STATus:OPERtion[:EVENt]?

Parameters

(None)

 

Returned Parameters

<NR1>

(Register Value)

Examples

STAT:OPER?

STATUS:OPERATIOBAL:EVENT?

Related Commands

*CLS STAT:OPER:NTR STAT:OPER:PTR

STAT:OPER:COND?

This query returns the value of the Operation Condition register. That is a read-only register which holds the real-time (unlatched) operational status of the power supply.

Query Syntax

STATus:OPERation:CONDition?

Parameters

(None)

Examples

STAT:OPER:COND? STATUS:OPERATION:CONDITION?

Returned Parameters

<NR1> (Register value)

Related Commands

(None)

Language Dictionary 41

Image 40
Contents PROGRAMMING GUIDE GPIB DC POWER SUPPLIES Agilent Part NoMicrofiche Part No JulySafety Guidelines Printing HistoryContents GENERAL INFORMATIONREMOTE PROGRAMMING LANGUAGE DICTIONARYDescription of Subsystem Commands ERROR MESSAGES STATUS REPORTINGSCPI CONFORMANCE INFORMATION COMPATIBILITY LANGUAGEGeneral Information About this GuideDocumentation Summary User’s GuidePrerequisites for Using this Guide VXIplug&play Power Product Instrument DriversDownloading and Installing the Driver Accessing Online HelpGPIB Capabilities Of The Power Supply Remote ProgrammingIntroduction To SCPI ConventionsTypes of SCPI Commands SCPI MessagesStructure of a SCPI Message Common CommandsParts of a SCPI Message Figure 2-1.Command Message StructureMessage Component VOLT LEV PROT CURRTraversing the Command Tree Query IndicatorMessage Unit Separator Root SpecifierFigure 2-2.Partial Command Tree Active Header PathThe Effect of Optional Headers Moving Among SubsystemsIncluding Common Commands SCPI QueriesValue Coupling SCPI Data FormatsExamples Listening FormatsTable 2-2.Suffixes and Multipliers ClassControlling the Output Disable the outputEnable the output Programming Voltage and CurrentWriting to the Display Saving and Recalling StatesProgramming Status Programming the Digital I/O Port System ConsiderationsThe GPIB Address A direct primary address and a secondary address DOS Drivers Agilent BASIC ControllersSample Program Code Error HandlingProgramming Some Power Supply Functions Controller Using Agilent 82335A InterfaceProgramming Some Power Supply Functions continued 22 Remote ProgrammingProgramming Some Power Supply Functions continued 24 Remote Programming Related Commands Common CommandsSubsystem Commands Language DictionaryDescription Of Common Commands Figure 3-1.Common Commands Syntax DiagramDescription Meaning and Type0 to ESR? IDN?Related Commands Query SyntaxOPC? OPT?Power-onStatus Clear Device Initialization PSC 0 *PSCMeaning and Type DescriptionMeaning and Type STB? Bit Configuration of Status Byte RegisterTST? Description of Subsystem Commands Calibration CommandsABOR Figure 3-2.Subsystem Commands Tree DiagramCurrent Subsystem CURR CURR TRIGCURR PROT STAT CURRENT:LEVEL 200 MADisplay Subsystem DIG DATADISP Digital I/O Port Programming ChartDISP MODE DISP TEXTDISP TEXT DEFAULT MODE enclosed in either single ‘ or double quotesInitiate Subsystem Measure SubsystemINIT INIT:CONT MEAS CURR? MEAS VOLT?Output Subsystem OUTPOUTP PROT CLE OUTP PROT DEL 0 orOUTP REL OUTP REL POLOUTP REL 1 OUTP REL OFF OUTP REL POL NORMStatus Subsystem STAT PRESStatus Operation Registers STAT OPER?STAT OPER NTR STAT OPER PTR STAT OPER ENABSTATUS OPERATION ENABLE? Status Questionable Registers STAT QUES?STAT:QUES:COND? STAT QUES ENABSystem Commands STAT QUES NTR STAT QUES PTRSYST ERR? corresponding Questionable Event registerTrigger Subsystem SYST LANGSYST VERS? TRIGVoltage Subsystem TRIG SOURVOLT VOLT TRIG VOLTAGE LEVEL 200 MVCommand Summary VOLT:PROTCommand Summary CommandCommand Parameters Characteristics in the Operating Guide Programming ParametersParameter Agilent Model and ValuePower Supply Status Structure Register CommandsStatus Reporting Operation Status GroupTable 4-2.Bit Configurations of Status Registers SignalFigure 4-1.Power Supply Status Model MeaningQuestionable Status Group Standard Event Status GroupTable 4-3.Status Questionable Commands CLS *ESR?Service Request Enable Register Initial Conditions At Power OnDetermining the Cause of a Service Interrupt Status Byte RegisterThe PON Power-OnBit Servicing an Operation Status Mode EventTable 4-4.Default Power On Register States Caused ByMonitoring Both Phases of a Status Transition Adding More Operation EventsServicing Questionable Status Events Table 4-5.Generating RQS from the CC EventSCPI Command Completion DFI Discrete Fault IndicatorRI Remote Inhibit Techniques in ANSI/IEEE StdError Messages Power Supply Hardware Error MessagesCalibration Error Messages System Error Messages60 Error Messages SCPI Confirmed Commands1 SCPI Approved CommandsSCPI Conformance Information SCPI VersionNON-SCPICommands1 Compatibility Language ARPS Command1 Table B-1.ARPS CommandsSimilar SCPI Table B-1.ARPS Commands continued Table B-1.ARPS Commands continued Index 68 Index Page Agilent Sales and Support Office United StatesLatin America Canada
Related manuals
Manual 70 pages 51.66 Kb Manual 142 pages 24.05 Kb

668xA, 669xA, 667xA, 664xA, 665xA specifications

Agilent Technologies has long been a pioneer in the production of high-performance electronic test and measurement instruments, particularly in the field of power sources. Among its notable offerings are the Agilent 667xA, 669xA, 665xA, 664xA, and 668xA series of power supplies. These instruments are designed to provide stable, reliable power for a variety of applications, including electronic testing, industrial processes, and research laboratories.

The Agilent 667xA series is characterized by its programmability and advanced measurement functions. These power supplies support a wide range of output voltages and currents, allowing for flexible configurations that cater to different testing needs. The built-in measurement capabilities enable users to monitor the voltage, current, and power with high precision, which is essential for ensuring optimal performance in electronic applications.

The Agilent 669xA series stands out with its high-power outputs, making it suitable for demanding applications. These power supplies deliver high voltage and current levels, making them ideal for testing high-performance devices, such as power amplifiers and motor drives. Additionally, the 669xA series includes features such as overvoltage protection and complex output sequencing to enhance the safety and reliability of the testing process.

The Agilent 665xA and 664xA series focus on delivering high accuracy and excellent regulation. These models are particularly known for their low noise operation, which is critical for sensitive applications where precision is paramount. The integrated programming capabilities allow users to automate testing sequences, thus improving efficiency in research and development settings.

The 668xA series features advanced digital signal processing that enhances the precision and stability of the output. Users benefit from features like remote sensing and monitoring, allowing feedback adjustments that maintain output accuracy despite cable losses. Furthermore, the 668xA models can integrate seamlessly with various test environments thanks to their LAN, GPIB, and USB connectivity options.

Overall, the Agilent 667xA, 669xA, 665xA, 664xA, and 668xA power supplies provide a comprehensive range of solutions for diverse electronic testing needs. With their advanced features, superb measurement capabilities, and robust performance, these instruments empower engineers and researchers to conduct their work with confidence, precision, and efficiency.