Invacare user manual Connection view of the IOC100P with CGA870

Page 11

SECTION 2—FEATURES

Selector Switch: Enables the user to select the desired setting, as well as “OFF” and “CF” (continuous flow). When not in use, the switch should be turned to the “OFF” position.

Oxygen Supply Outlet: Use this fitting to attach a standard cannula.

Cylinder Alignment Pins: When assembling the unit, these parallel pins MUST be inserted into the holes on the CGA870 post valve.

CGA870 Seal Washer [Metal Washer with Viton® or equivalent Seal]: Creates the interface between the post valve and the conserver regulator. Besides offering a rugged interface, it also surrounds the oxygen path in a ring of stainless steel or brass.

Vent Hole: Maintains proper internal pressure. DO NOT obstruct with any object, such as a label or tight‐fitting carrying bag.

NOTE: Use only a CGA870 Seal Washer [Metal Washer with Viton Seal]. DO NOT use a plastic compression gasket.

FIGURE 2.2 Connection view of the IOC100P with CGA870

Connection

Part No 1160840

11

Invacare® Element™

Image 11
Contents Invacare Element Pneumatic Conserver Accessories Warning Table of Contents Word Special NotesPrinciples of Operation Introduction Operating Information General GuidelinesGeneral Guidelines General Guidelines Model IOC100P FeaturesConnection view of the IOC100P with CGA870 Installing the Conserving Regulator Cylinder post valve Operating Instructions Cannula Placement Operating Instructions Selector Switch Removing the Conserving Regulator Regulator Typical Product ParametersOxygen Cylinder Duration Maintenance Solution TroubleshootingLimited Warranty USA