SECTION 8—TROUBLESHOOTING

SECTION 8—TROUBLESHOOTING

If oxygen conserving regulator fails to function, consult the Troubleshooting guide. If problem cannot be corrected, consult your Dealer/Home Care equipment provider.

PROBLEM

POSSIBLE CAUSE

SOLUTION

 

 

 

No flow.

1. Cylinder valve is closed.

1. Turn the cylinder valve to

Unit does not pulse.

2. Regulator in “Off”

the “ON” position.

 

2. Set to prescribed setting.

 

position.

 

3. Cylinder empty.

3. Replace cylinder.

 

4. Oxygen conserving

4. Check position of

 

regulator not sensing

cannula in nose. DO

 

breath.

NOT breathe through

 

 

mouth.

 

5. Oxygen cannula is

5. Remove kinks. Clean or

 

blocked or kinked.

replace, if necessary.

 

 

 

Unit flows or pulses

1. Unit is set to the “CF”

1. Turn the selector switch

continuously.

position.

to the appropriate

 

 

delivery setting.

 

2. Unit was not set to

2. Turn the selector switch

 

to “OFF,” wait a few

 

“OFF” prior to opening

 

moments, then set at

 

the cylinder valve.

 

proper delivery setting.

 

 

 

 

3. Remove obstructions,

 

3. Vent hole is obstructed.

such as labels or carrying

 

 

bag, and resume use as

 

 

usual.

 

 

 

Replacement Parts - Accessories

HF2POST6BAG Carrying Bag or HF2POST9BAG Carrying Bag

Invacare® Element™

22

Part No 1160840

Image 22
Contents Invacare Element Pneumatic Conserver Accessories Warning Table of Contents Special Notes WordPrinciples of Operation Introduction General Guidelines Operating InformationGeneral Guidelines General Guidelines Features Model IOC100PConnection view of the IOC100P with CGA870 Installing the Conserving Regulator Cylinder post valve Operating Instructions Cannula Placement Operating Instructions Selector Switch Removing the Conserving Regulator Typical Product Parameters RegulatorOxygen Cylinder Duration Maintenance Troubleshooting SolutionLimited Warranty USA